一种新的化学发光免疫分析法在测量抗β2 糖蛋白 1 和抗心磷脂抗体中的临床性能。

The Clinical Performance of a New Chemiluminescent Immunoassay in Measuring Anti-β2 Glycoprotein 1 and Anti-Cardiolipin Antibodies.

机构信息

Department of Laboratory Medicine, Peking University Third Hospital, Beijing, China (mainland).

Research Center of Clinical Epidemiology, Peking University Third Hospital, Beijing, China (mainland).

出版信息

Med Sci Monit. 2018 Sep 26;24:6816-6822. doi: 10.12659/MSM.910369.

Abstract

BACKGROUND Laboratory criterion is needed for the classification of antiphospholipid syndrome (APS), which contain anticardiolipin antibodies (aCL) and anti-β2-glycoprotein 1 antibodies (aβ2GP1). They are commonly identified by enzyme-linked immunosorbent assay (ELISA), but lack standardized kits, resulting in substantial variations in the antibody positivity between different laboratories. The emergence of chemiluminescence automated -BIO-FLASH may improve the situation. MATERIAL AND METHODS We selected 185 patients with APS, systemic lupus erythematosus (SLE), infertility, connective tissue disease (CTD), and other conditions in Peking University Third Hospital. We tested the aCL and aβ2GP1 levels by EUROIMMUN ELISA and 105 patients had at least one positive result for aCL and aβ2GP1, while the others had negative results. We retested them by chemiluminescence assay (CIA) and analyzed the result and compared the coincidence rate. The IgM levels were retested by AESKU ELISA. Data were analyzed using SPSS. RESULTS Our result suggested that CIA had good performance for IgG isotype of aCL and aβ2GP1 in the coincidence rate. The positive coincidence rate of aCL IgM between CIA and EUROIMMUN ELISA was only 41.67%, but two ELISA kits showed good coincidence, CIA and AESKU ELISA had an obviously higher positive rate. CIA and AESKU had a higher coincidence than that of AESKU and EUROIMMUN in aβ2GP1-IgM. CONCLUSIONS The new automated CIA BIO-FLASH is suitable for detecting aCL and aβ2GP1 antibodies, especially IgG isotype, which may provide an alternative to time-consuming conventional ELISA method.

摘要

背景

抗磷脂综合征(APS)的分类需要实验室标准,其中包含抗心磷脂抗体(aCL)和抗-β2-糖蛋白 1 抗体(aβ2GP1)。它们通常通过酶联免疫吸附试验(ELISA)进行鉴定,但缺乏标准化试剂盒,导致不同实验室之间抗体阳性率存在很大差异。化学发光自动化-BIO-FLASH 的出现可能会改善这种情况。

材料和方法

我们选择了北京大学第三医院的 185 例 APS、系统性红斑狼疮(SLE)、不孕、结缔组织病(CTD)和其他疾病患者。我们使用 EUROIMMUN ELISA 检测 aCL 和 aβ2GP1 水平,其中 105 例患者至少有一种 aCL 和 aβ2GP1 阳性结果,而其他患者则为阴性。我们用化学发光法(CIA)对其进行了复测,并分析了结果并比较了符合率。我们用 AESKU ELISA 复测了 IgM 水平。使用 SPSS 进行数据分析。

结果

我们的结果表明,CIA 在符合率方面对 aCL 和 aβ2GP1 的 IgG 同种型具有良好的性能。CIA 和 EUROIMMUN ELISA 检测 aCL IgM 的阳性符合率仅为 41.67%,但两种 ELISA 试剂盒具有良好的符合率,CIA 和 AESKU ELISA 的阳性率明显较高。CIA 和 AESKU 在 aβ2GP1-IgM 方面的符合率高于 AESKU 和 EUROIMMUN。

结论

新的自动化 CIA BIO-FLASH 适用于检测 aCL 和 aβ2GP1 抗体,特别是 IgG 同种型,它可能为耗时的传统 ELISA 方法提供替代方案。