Division of Neonatal-Perinatal Medicine, Department of Pediatrics, Saint Louis University School of Medicine, St. Louis, Missouri, USA.
Division of Neonatal-Perinatal Medicine, Department of Pediatrics, Saint Louis University School of Medicine, St. Louis, Missouri,
Neonatology. 2019;115(1):28-35. doi: 10.1159/000493156. Epub 2018 Oct 2.
Bubble continuous positive airway pressure (bCPAP) generates a variable pressure due to bubble size and frequency, and these pressure oscillations may contribute to lung recruitment and gas exchange. Single-outlet bCPAP devices generate larger pressure fluctuation than bCPAP devices with diffusers, but clinical efficacy is unclear. Our hypothesis was that a single-outlet bCPAP device (B&B Bubbler©) would decrease extubation failure and improve oxygenation compared with a bCPAP device with a diffuser (BabiPlus©).
Bench testing of bCPAP devices with the RAM© cannula determined pressure averages, fluctuations, and CPAP transmission at prongs. The B&B Bubbler© and the BabiPlus© devices were compared for oxygenation and bCPAP failure within 72 h of extubation in a randomized trial in infants less than 1,250 g (n = 43). Additionally, a cohort study with infants extubated to BabiPlus© (n = 128), B&B Bubbler© (n = 119), and noninvasive positive pressure ventilation (NIPPV) (n = 107) were compared.
Small differences in pressure fluctuations between bCPAP devices are attenuated by the RAM© cannula. Pressures at RAM© prongs are 62% of set pressure. The bCPAP extubation failure rate (35%) and oxygenation were similar between the bCPAP devices. bCPAP success was associated with lower FiO2 at extubation and decreased need for additional FiO2 in both groups. In the setting of the RAM© cannula, there were no differences in extubation failure rate for infants extubated to bCPAP or NIPPV.
There were no clinical differences at extubation between the B&B Bubbler© and BabiPlus© devices. Oxygenation may predict bCPAP failure. The use of the RAM© cannula as a bCPAP interface should be more closely examined.
由于气泡大小和频率的原因,气泡持续气道正压通气(bCPAP)会产生可变压力,这些压力波动可能有助于肺复张和气体交换。单出口 bCPAP 设备产生的压力波动比带有扩散器的 bCPAP 设备大,但临床疗效尚不清楚。我们的假设是,与带有扩散器的 bCPAP 设备(BabiPlus©)相比,单出口 bCPAP 设备(B&B Bubbler©)将降低拔管失败率并改善氧合。
使用 RAM© 套管对 bCPAP 设备进行 bench 测试,以确定压力平均值、波动和鼻塞 CPAP 传输。在一项小于 1250g 婴儿的随机试验中,在拔管后 72 小时内比较 B&B Bubbler©和 BabiPlus©设备的氧合和 bCPAP 失败情况(n=43)。此外,还对拔管至 BabiPlus©(n=128)、B&B Bubbler©(n=119)和无创正压通气(NIPPV)(n=107)的婴儿进行了队列研究。
RAM©套管会减弱 bCPAP 设备之间压力波动的微小差异。RAM©鼻塞的压力为设定压力的 62%。bCPAP 拔管失败率(35%)和氧合在两种设备之间相似。bCPAP 成功与拔管时较低的 FiO2 和两组中需要额外 FiO2 的减少有关。在 RAM©套管的情况下,拔管至 bCPAP 或 NIPPV 的婴儿的拔管失败率没有差异。
B&B Bubbler©和 BabiPlus©设备在拔管时没有临床差异。氧合可能预测 bCPAP 失败。应更仔细地检查 RAM©套管作为 bCPAP 接口的使用。