Demoulin Bruno, Ioan Iulia, Duvivier Claude, Bonabel Claude, Schweitzer Cyril, Marchal François, Demoulin-Alexikova Silvia
EA 3450 DevAH, Department of Physiology, Faculty of Medicine, University of Lorraine, Nancy, France.
Department of Pediatric Respiratory Function Testing, Children's Hospital, Vandoeuvre, France.
Front Physiol. 2018 Oct 9;9:1408. doi: 10.3389/fphys.2018.01408. eCollection 2018.
Multi-center studies in specific airway resistance have shown significant inter laboratory variability. Comparison of plethysmographic equipment using a lung model easily transportable from one site to another should be of help to international normative studies. A resistor made of parallel capillary tubes - insuring adequate linearity within 1 L/sec - was connected to a glass bottle. Thermal time constants were measured while the bottle was empty and while stuffed with steel wool. In the latter, isothermal condition was estimated to occur only at very low frequency (around 0.01 Hz) and gas compression was polytropic up to 0.6 Hz. With the empty analog, adiabatic gas compression was estimated to occur at frequencies ≥0.2 Hz, and more accurate volume estimation was obtained. The empty analog volume and specific resistance measured in a body plethysmograph on different days indicated within 5% accuracy as well as intersession repeatability. It is concluded that a physical analog built out of simple material provides accurate measurements of specific resistance. The apparatus should be of help to compare plethysmographic equipments from different laboratories.
多中心关于比气道阻力的研究显示实验室间存在显著差异。使用易于从一个地点运输到另一个地点的肺模型对比体积描记设备,应有助于国际规范研究。由平行毛细管制成的电阻器(确保在1升/秒内有足够的线性度)连接到一个玻璃瓶上。在瓶子为空和装满钢丝绒时测量热时间常数。在后者情况下,估计仅在非常低的频率(约0.01赫兹)下才会出现等温条件,并且在高达0.6赫兹时气体压缩是多方的。对于空的模拟装置,估计在频率≥0.2赫兹时会发生绝热气体压缩,并且获得了更准确的体积估计。在不同日期在体容积描记仪中测量的空模拟装置体积和比电阻显示精度在5%以内,以及不同测量时段间的可重复性。得出的结论是,由简单材料制成的物理模拟装置可提供比电阻的准确测量。该设备应有助于比较不同实验室的体积描记设备。