Beeksma Brad, Lehmann Joerg
1 Departmentof Radiation Oncology, Calvary Mater Newcastle , Newcastle, NSW , Australia.
2 Institute of Medical Physics, School of Physics, University of Sydney , Sydney, NSW , Australia.
Br J Radiol. 2019 Feb;92(1094):20180743. doi: 10.1259/bjr.20180743. Epub 2018 Nov 15.
: The objective of this work is to characterise out- of- field leakage radiation emanating from clinical applicators of the WOmed T-200 kilovoltage therapy machine.
: To identify points of leakage, radiosensitive film was affixed to the walls and base plate of each applicator. Quantitative assessment of leakage radiation was conducted with a 0.23 cm ionisation chamber following the International Electrotechnical Commission standard. Film was also used to illustrate leakage distribution in the patient plane. Angular and energy dependences of the leakage radiation were quantified as well as a two-dimensional leakage profile in the plane parallel to one applicator.
: Leakage was found when the diameter of primary collimator of the kV tube exceeded the external dimension of the applicator wall. In the patient plane all applicators showed similar leakage rates with the leakage distribution dependent upon applicator design. Mean patient plane leakage was 1.37% of central axis air kerma rate, exceeding the 0.5% limit specified in the standard. Leakage was shown to be profoundly energy dependent with maximum leakage of 11.8% for the 200 kV beam, 1.3% for 150 kV and 0.2% for 100 kV. Angular dependence measurements showed a 10.3% change in leakage between the minimum and maximum positions.
The combination of shielding thickness, primary collimator design and applicator dimensions permits unwanted radiation to contribute dose outside the treatment field when energies ≥150 kV are used.
: Even carefully designed modern kv therapy systems can exhibit leakage in some areas. Thorough assessment of leakage is needed prior to release for clinical use.
本研究旨在对WOmed T - 200千伏治疗机临床施源器发出的射野外漏辐射进行特征描述。
为确定漏射点,将放射敏感胶片贴于每个施源器的壁和底板上。按照国际电工委员会标准,使用0.23厘米电离室对漏辐射进行定量评估。胶片还用于说明患者平面内的漏射分布情况。对漏辐射的角度和能量依赖性进行了量化,并给出了与一个施源器平行平面内的二维漏射分布图。
当千伏管的初级准直器直径超过施源器壁的外部尺寸时,会出现漏射。在患者平面内,所有施源器的漏射率相似,漏射分布取决于施源器设计。患者平面内的平均漏射率为中心轴空气比释动能率的1.37%,超过了标准规定的0.5%的限值。结果表明,漏射与能量密切相关,200千伏射线束的最大漏射率为11.8%,150千伏时为1.3%,100千伏时为0.2%。角度依赖性测量显示,最小和最大位置之间的漏射变化为10.3%。
当使用能量≥150千伏时,屏蔽厚度、初级准直器设计和施源器尺寸的组合会使非预期辐射在治疗野外产生剂量。
即使是精心设计的现代千伏治疗系统在某些区域也可能出现漏射。在临床使用前,需要对漏射进行全面评估。