通过超声扫描测定硅胶晶状体植入术后假性眼轴长度增加情况。

Pseudo axial length increase after silicone lens implantation as determined by ultrasonic scans.

作者信息

Milauskas A T, Marney S

出版信息

J Cataract Refract Surg. 1988 Jul;14(4):400-2. doi: 10.1016/s0886-3350(88)80147-0.

Abstract

Repeat A-scans of 45 patients who had silicone intraocular lens (IOL) implantation indicated an apparent increase in the axial length of the eyes. The average increase of 1.045 mm over the original axial length was noted when using the pseudophakic mode for the axial length determination. The actual velocity of sound in eyes with a silicone IOL is 1,486 M/sec compared to 1,548 M/sec in an eye with a polymethylmethacrylate implant. This finding is clinically significant when one is comparing the axial length of two eyes, one of which is pseudophakic. A 1 mm difference in axial length can cause serious doubts about the validity of one or the other of the scans if the type of implant is not known. The velocity of RMX-3 silicone was determined to be 1,049 M/sec.

摘要

对45例行硅酮人工晶状体(IOL)植入术的患者进行重复A超扫描显示,眼轴长度明显增加。使用假晶状体模式测定眼轴长度时,发现平均眼轴长度比原始眼轴长度增加了1.045mm。与植入聚甲基丙烯酸甲酯人工晶状体的眼睛相比,植入硅酮人工晶状体的眼睛中实际声速为1486米/秒,而前者为1548米/秒。当比较两只眼睛(其中一只为假晶状体眼)的眼轴长度时,这一发现具有临床意义。如果不知道植入物的类型,眼轴长度相差1mm可能会对其中一次扫描结果的有效性产生严重怀疑。已确定RMX - 3硅酮的声速为1049米/秒。