Zhao Zhichun, Ivanov Kamen, Lubich Ludwig, Omisore Olatunji Mumini, Mei Zhanyong, Fu Nan, Chen Jinying, Wang Lei
Annu Int Conf IEEE Eng Med Biol Soc. 2018 Jul;2018:4965-4968. doi: 10.1109/EMBC.2018.8513457.
Recent advancement in technology has brought about increase in the application areas of wearable electroencephalographic devices. In that, new types of electrodes take place, and particular attention is needed to ensure the required quality of obtained signals. In this study, we evaluate electrode-skin impedance and signal quality for several kinds of electrodes when used in conditions typical for wearable devices. Results suggest that active dry electrode coated with gold alloy is superior while it was challenging to obtain appropriate signal quality when using passive dry electrodes. We also demonstrate electrode-skin impedance measurement using the analog frontend ADS1299, which is suitable for implementation in wearable devices.
近年来技术的进步使得可穿戴脑电图设备的应用领域不断扩大。在此过程中,新型电极不断涌现,因此需要特别注意确保所获取信号的质量。在本研究中,我们评估了几种电极在可穿戴设备典型使用条件下的电极-皮肤阻抗和信号质量。结果表明,涂有金合金的有源干电极表现更优,而使用无源干电极时要获得合适的信号质量则颇具挑战。我们还展示了使用适用于可穿戴设备的模拟前端ADS1299进行电极-皮肤阻抗测量的方法。