Swiontkowski M F, Schlehr F, Collins J C, Sanders R, Pou A
Department of Orthopaedics and Rehabilitation, Vanderbilt University, Nashville, Tennessee 37232.
Calcif Tissue Int. 1988 Aug;43(2):103-7. doi: 10.1007/BF02555155.
The development of laser Doppler flowmetry (LDF) has provided a real-time, reliable method for monitoring capillary perfusion in multiple tissues. LDF has potential for the experimental and clinical assessment of bone blood flow. To compare the accuracy and ease of use of two commercially available laser Doppler flowmeters with different mechanisms for processing the Doppler-shifted component of light, estimates of bone blood flow were obtained in a sheep model using the two systems, and the values derived then compared with estimates of bone blood flow also obtained in a sheep using the technique of injection of labeled microspheres. The single-channel laser Doppler flowmeter, the LD 5000, processes the reference and Doppler-shifted beams on the surface of a single photodetector using optical heterodyning for measurement. The dual-channel flowmeter, the Periflux 2, uses two optical fibers to transmit reference and Doppler-shifted light to two separate photodetectors. The differential amplification and detection system improves the signal-to-noise ratio. Measurement of both metaphyseal (cancellous) and diaphyseal (cortical) blood flow using both LDF systems was compared with values obtained with an injection of 85Sr-labeled microspheres in three sheep. The LDF measurements were repeated after occlusion of the left femoral artery, and a 46Sc microsphere injection was performed prior to animal sacrifice. Two of the animals developed vasomotor instability, resulting in poor correlation between the measurements obtained with the Periflux 2, which is motion sensitive, and the values obtained with the microsphere method. High correlation was apparent in two of the three animals for the LD 5000 and the microsphere values.(ABSTRACT TRUNCATED AT 250 WORDS)
激光多普勒血流仪(LDF)的发展为监测多种组织中的毛细血管灌注提供了一种实时、可靠的方法。LDF在骨血流的实验和临床评估方面具有潜力。为比较两种具有不同光多普勒频移分量处理机制的市售激光多普勒血流仪的准确性和易用性,在绵羊模型中使用这两种系统获得骨血流估计值,并将所得值与同样在绵羊中使用注射标记微球技术获得的骨血流估计值进行比较。单通道激光多普勒血流仪LD 5000利用光学外差法在单个光电探测器表面处理参考光束和多普勒频移光束以进行测量。双通道血流仪Periflux 2使用两根光纤将参考光和多普勒频移光传输到两个独立的光电探测器。差分放大和检测系统提高了信噪比。在三只绵羊中,将使用两种LDF系统测量的干骺端(松质骨)和骨干(皮质骨)血流值与注射85Sr标记微球获得的值进行比较。在阻断左股动脉后重复进行LDF测量,并在动物处死前进行46Sc微球注射。其中两只动物出现血管舒缩不稳定,导致对运动敏感的Periflux 2测量值与微球法获得的值之间相关性较差。在三只动物中的两只中,LD 5000测量值与微球值之间存在明显的高相关性。(摘要截短于250字)