Suppr超能文献

扩瞳对视荧光寿命成像检查的影响。

Impact of mydriasis in fluorescence lifetime imaging ophthalmoscopy.

机构信息

Department of Ophthalmology, David Geffen School of Medicine at UCLA, Doheny Eye Center, Los Angeles, CA.

Department of Ophthalmology, David Geffen School of Medicine at UCLA, Stein Eye Institute, Los Angeles, CA.

出版信息

PLoS One. 2018 Dec 28;13(12):e0209194. doi: 10.1371/journal.pone.0209194. eCollection 2018.

Abstract

Fluorescence lifetime imaging ophthalmoscopy (FLIO) is a novel technique that measures in vivo autofluorescence intensity decay over time of endogenous fluorophores in the retina. The Heidelberg Engineering FLIO system was used to obtain two 30 degree scans centered on the fovea of both eyes. The FLIO system uses a 473nm blue scanning laser light source and the emitted fluorescence is detected in two wavelengths channels, short and long spectral channels (SSC, LSC). Since the mydriatic status influence the FLIO result, the impact of mydriasis on FLIO need to be clarified. In this prospective, observational study, the impact of mydriasis on measurements from fluorescence lifetime imaging ophthalmoscope (FLIO) images in normal subjects were evaluated. 12 healthy participants (24 eyes) were volunteered and all subjects were scanned twice and the mean fluorescence lifetime (τm) values were computed with dilation and without dilation on different days. Intraclass correlation coefficients (ICC) and coefficients of variation (CV) were calculated from the measured τm in dilated, nondilated and between the dilated and non-dilated setting. Test duration was also compared and correlated with lifetimes in both settings. Repeatability was excellent for both the dilation and non-dilation settings (ICC; 0.967-0.996; 0.926-0.986, respectively). The agreement between the dilation and non-dilation settings, however, were lower (ICC; 0.688-0.970). The τm in the non-dilation setting was significantly longer than in the dilation setting for the SSC (P<0.05). The FLIO test duration in the non-dilation setting was significantly longer than with dilation for the SSC (P <0.05). Although good repeatability in τm measurements between imaging sessions were observed both with and without dilation, the agreement was not as good when comparing dilated with non-dilated measurements. Since FLIO without mydriasis results in longer τm in the SSC and takes a longer time for image acquisition, maximal dilation is recommended for FLIO testing.

摘要

荧光寿命成像检眼镜(FLIO)是一种新型技术,可测量视网膜内源性荧光团的体内自发荧光强度随时间的衰减。使用海德堡工程公司的 FLIO 系统获得了以双眼黄斑为中心的两个 30 度扫描。FLIO 系统使用 473nm 蓝色扫描激光光源,发射的荧光在两个波长通道(短和长光谱通道(SSC,LSC))中检测。由于散瞳状态会影响 FLIO 结果,因此需要阐明散瞳对 FLIO 的影响。在这项前瞻性观察研究中,评估了散瞳对正常受试者荧光寿命成像检眼镜(FLIO)图像测量的影响。12 名健康志愿者(24 只眼)自愿参与研究,所有受试者在不同日期分别进行散瞳和非散瞳扫描两次,并计算平均荧光寿命(τm)值。在散瞳和非散瞳状态下测量的 τm 值计算出组内相关系数(ICC)和变异系数(CV)。还比较了测试时间并将其与两种状态下的寿命相关联。散瞳和非散瞳设置的重复性均很好(ICC;0.967-0.996;0.926-0.986)。然而,散瞳和非散瞳设置之间的一致性较低(ICC;0.688-0.970)。非散瞳设置中的 τm 明显长于散瞳设置中的 τm(SSC,P<0.05)。非散瞳设置中的 FLIO 测试时间明显长于散瞳设置中的 SSC(P<0.05)。尽管在散瞳和非散瞳成像过程中都观察到 τm 测量的重复性很好,但在比较散瞳和非散瞳测量值时,一致性却不佳。由于非散瞳的 FLIO 会导致 SSC 中的 τm 更长,并且图像采集时间更长,因此建议对 FLIO 测试进行最大程度散瞳。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/9407/6310355/1ea0e89fe3cf/pone.0209194.g001.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验