Mutwalli Hussam, Braian Michael, Mahmood Deyar, Larsson Christel
Department of Prosthodontics, Faculty of Odontology, Malmö University, Smedjegatan 16, 214 21 Malmö, Sweden.
Department of Materials Science and Technology, Faculty of Odontology, Malmö University, Smedjegatan 16, 214 21 Malmö, Sweden.
Int J Dent. 2018 Nov 26;2018:5189761. doi: 10.1155/2018/5189761. eCollection 2018.
To measure the trueness and precision under repeatable conditions for different intraoral scanners (IOSs) when scanning fully edentulous arch with multiple implants.
Three IOSs and one industrial scanner were used to scan one edentulous master cast containing five implant scan bodies and three spheres. The cast was scanned thirty times with each scanner device. All scans were analyzed in the inspect software, and three-dimensional locations of the implants and the interarch distance between the spheres were measured. The values were compared to measurements made with one coordinate measuring machine (true value). One-way ANOVA was used to calculate the differences between IOSs and in comparison with the true value.
Significant differences were found between all IOSs. For the implant measurements, Trios 3 had the lowest trueness (≤114 m), followed by Trios 3 mono (≤63 m) and Itero element (≤-41 m). Trios had the lowest precision (≤135 m), followed by Itero element (≤101 m) and Trios 3 mono (≤100 m). With regard to the interarch distance measurements, Trios 3 had the lowest trueness (≤68 m), followed by Trios 3 mono (≤45 m) and Itero element (≤40 m). Trios 3 had the lowest precision (≤206 m), followed by Itero element (≤124 m) and Trios 3 mono (≤111 m).
The results from this in vitro study suggest that precision is low for the tested IOS devices when scanning fully edentulous arches with multiple implants.
测量不同口腔内扫描仪(IOS)在对带有多个种植体的全口无牙颌弓进行扫描时,在可重复条件下的准确性和精密度。
使用三台IOS和一台工业扫描仪对一个含有五个种植体扫描体和三个球体的无牙颌主模型进行扫描。每个扫描仪设备对该模型进行30次扫描。所有扫描图像均在检测软件中进行分析,并测量种植体的三维位置以及球体之间的颌间距离。将这些测量值与使用一台坐标测量机所测得的(真实值)进行比较。采用单因素方差分析来计算IOS之间以及与真实值相比的差异。
在所有IOS之间均发现了显著差异。对于种植体测量,Trios 3的准确性最低(≤114μm),其次是Trios 3 mono(≤63μm)和Itero element(≤-41μm)。Trios的精密度最低(≤135μm),其次是Itero element(≤101μm)和Trios 3 mono(≤100μm)。关于颌间距离测量,Trios 3的准确性最低(≤68μm),其次是Trios 3 mono(≤45μm)和Itero element(≤40μm)。Trios 3的精密度最低(≤206μm),其次是Itero element(≤124μm)和Trios 3 mono(≤111μm)。
这项体外研究的结果表明,当对带有多个种植体的全口无牙颌弓进行扫描时,所测试的IOS设备的精密度较低。