Abstract

PURPOSE

To evaluate a custom-made ocular fluorometer for detection of intensity of light scatter (ILS) from the anterior chamber (A/C) as an objective measure of aqueous flare.

METHODS

The fluorometer, equipped with a lock-in amplifier, was employed in the scatter mode to detect ILS from A/C. Measurements were performed with two illumination slit widths of 0.5 and 0.25 mm. The axial resolution at these slit widths were 80 and 200 μm, respectively. Healthy and pseudophakic eyes, with grade 0 Standardization of Uveitis Nomenclature (SUN) score, were employed as control subjects. ILS was also recorded in a cohort of patients who had undergone phacoemulsification and showed grades 1+ or 2+ on postoperative days 1 and 4.

RESULTS

The inter- and intraobserver variabilities in the measurement of ILS were not significant. In cataract patients, ILS was significantly higher on postoperative day 1 relative to healthy eyes. By day 4, ILS decreased significantly and was only marginally different from ILS in quiet pseudophakic eyes or healthy eyes. Eyes with higher SUN scores showed proportionately increased ILS. The receiver-operator characteristic analysis indicated no advantage in using the smaller slit width in discriminating ILS at different SUN scores although it provided higher axial resolution.

CONCLUSIONS

The lock-in-based spot fluorometer is reliable for measurement of ILS with high precision and accuracy.The measured ILS correlates linearly with SUN scores and can be used to provide a higher granularity for recording aqueous flare.

TRANSLATIONAL RELEVANCE

The instrument can be used in the clinical management of uveitis and drug development toward uveitis.

摘要

目的

评估一种定制的眼用荧光计,用于检测前房(A/C)的光散射强度(ILS),作为房水闪光的客观测量指标。

方法

配备锁相放大器的荧光计在散射模式下用于检测前房的ILS。测量采用0.5和0.25mm两种照明狭缝宽度进行。这些狭缝宽度下的轴向分辨率分别为80和200μm。将葡萄膜炎命名标准化(SUN)评分为0级的健康眼和人工晶状体眼作为对照对象。还记录了一组接受超声乳化手术且术后第1天和第4天SUN评分为1+或2+的患者的ILS。

结果

观察者间和观察者内测量ILS的变异性均不显著。在白内障患者中,术后第1天的ILS显著高于健康眼。到第4天,ILS显著下降,与安静的人工晶状体眼或健康眼中的ILS仅略有差异。SUN评分较高的眼,ILS相应增加。受试者工作特征分析表明,尽管较小的狭缝宽度提供了更高的轴向分辨率,但在区分不同SUN评分的ILS时并无优势。

结论

基于锁相的点荧光计在高精度和准确性测量ILS方面是可靠的。测得的ILS与SUN评分呈线性相关,可用于提供更高粒度的房水闪光记录。

转化相关性

该仪器可用于葡萄膜炎的临床管理和葡萄膜炎药物研发。