Arba Mosquera Samuel, de Ortueta Diego
Grupo de Investigación de Cirugía Refractiva y Calidad de Visión, Instituto de Oftalmobiología Aplicada, University of Valladolid, Valladolid, Spain
SCHWIND eye-tech-solutions GmbH, Kleinostheim, Germany
J Refract Surg. 2011 Feb 1;27(2):148-152. doi: 10.3928/1081597X-20100224-01.
To report a case in which a tissue saving program in an aberrated eye was used.
A new algorithm for the selection of an optimized set of Zernike terms in customized treatments for laser corneal refractive surgery was developed and clinically tested. Ablation was performed using the SCHWIND ESIRIS excimer laser. Pre- and postoperative corneal wave aberrations were analyzed using the Keratron Scout videokeratoscope (Optikon 2000).
Required ablation was reduced by approximately 15% compared to full customized correction. Refraction was corrected to subclinical levels, uncorrected distance visual acuity improved to 20/20, corrected distance visual acuity gained 2 lines, aberrations were reduced by approximately 40% compared to preoperative baseline levels, and the functional optical zone of the cornea was enlarged by approximately 40% compared to preoperative baseline levels. Trefoil, coma, spherical aberration, and the root-mean-square value of the higher order aberrations were reduced.
Eliminating all higher order aberrations may not optimize visual function in highly aberrated eyes. The new algorithm minimized tissue removal in refractive surgery but further clinical evaluations are required to confi rm preliminary results.
报告一个在异常眼中使用组织保存程序的病例。
开发了一种新算法,用于在激光角膜屈光手术的定制治疗中选择优化的一组泽尼克项,并进行了临床测试。使用SCHWIND ESIRIS准分子激光进行消融。使用Keratron Scout角膜地形图仪(Optikon 2000)分析术前和术后的角膜波像差。
与完全定制矫正相比,所需消融减少了约15%。屈光不正矫正至亚临床水平,未矫正远视力提高到20/20,矫正远视力提高了2行,像差与术前基线水平相比减少了约40%,角膜的功能光学区与术前基线水平相比扩大了约40%。三叶草像差、慧差、球差和高阶像差的均方根值均降低。
消除所有高阶像差可能无法使高度异常眼的视觉功能达到最佳。新算法在屈光手术中减少了组织切除,但需要进一步的临床评估来证实初步结果。