To compare the clinical outcomes of myopiacorrected with corneal-wavefront-guided (CWG) laser in situ keratomileusis (LASIK) with AMARIS 1050S (SCHWIND eye-tech-solutions GmbH & Co. KG) and corneal-topography-guided (CTG) LASIK with WaveLight EX500 (Alcon Laboratories, Fort Worth, TX).
In this prospective, pseudo-randomized expanded cohort study, a total of 266 patients were subjected to binocular LASIK surgery, either with WaveLight EX500 (WaveLight group) or Amaris 1050S (AMARIS group) platforms. Data related to right eyes were selected for analysis. Corneal higher-order aberration (HOA) was selected as the primary endpoint; while visual acuity and refraction were the secondary endpoints. All the endpoints were assessed at 3 months postoperatively.
There were 134 eyes in the AMARIS group and 132 eyes in the WaveLight group. After 3 months of postoperative follow-up, spherical and coma aberrations were significantly lower (P < 0.05) in the WaveLight group (spherical aberration: - 0.104 ± 0.199 µm; coma aberration: - 0.117 ± 0.202 µm) in comparison with the AMARIS group (spherical aberrations: 0.254 ± 0.146 µm; coma aberrations: 0.316 ± 0.297 µm). In the AMARIS group, 96.3% of the eyes achieved an uncorrected distance visual acuity (UDVA) of 20/20 while in the WaveLight group, 96.2% of the eyes achieved an UDVA of 20/20. Furthermore, the mean postoperative manifest refraction spherical equivalent (MRSE) was - 0.02 ± 0.28 in the AMARIS group and - 0.05 ± 0.21 in the WaveLight group (P = 0.34).
Both WaveLight EX500 and Amaris 1050S LASIK showed excellent refractive and visual outcomes. In addition, the WaveLight group showed minimal spherical and coma aberrations when compared to the AMARIS group.