Suppr超能文献

使用两种不同的术中三维成像导航系统进行脊柱手术时的辐射剂量与图像质量比较。

Radiation dose and image quality comparison during spine surgery with two different, intraoperative 3D imaging navigation systems.

作者信息

Nachabe Rami, Strauss Keith, Schueler Beth, Bydon Mohamad

机构信息

Image Guided Therapy Systems, Philips Healthcare, Best, The Netherlands.

Department of Radiology, Cincinnati Children's Hospital, Cincinnati, OH, USA.

出版信息

J Appl Clin Med Phys. 2019 Feb;20(2):136-145. doi: 10.1002/acm2.12534. Epub 2019 Jan 24.

Abstract

Careful protocol selection is required during intraoperative three-dimensional (3D) imaging for spine surgery to manage patient radiation dose and achieve clinical image quality. Radiation dose and image quality of a Medtronic O-arm commonly used during spine surgery, and a Philips hybrid operating room equipped with XperCT C-arm 3D cone-beam CT (hCBCT) are compared. The mobile O-arm (mCBCT) offers three different radiation dose settings (low, standard, and high), for four different patient sizes (small, medium, large, and extra large). The patient's radiation dose rate is constant during the entire 3D scan. In contrast, C-CBCT spine imaging uses three different field of views (27, 37, and 48 cm) using automatic exposure control (AEC) that modulates the patient's radiation dose rate during the 3D scan based on changing patient thickness. hCBCT uses additional x-ray beam filtration. Small, medium, and large trunk phantoms designed to mimic spine and soft tissue were imaged to assess radiation dose and image quality of the two systems. The estimated measured "patient" dose for the small, medium, and large phantoms imaged by the mCBCT considering all the dose settings ranged from 9.4-27.6 mGy, 8.9-33.3 mGy, and 13.8-40.6 mGy, respectively. The "patient" dose values for the same phantoms imaged with hCBCT were 2.8-4.6 mGy, 5.7-10.0 mGy, and 11.0-15.2 mGy. The CNR for the small, medium, and large phantoms was 2.9 to 3.7, 2.0 to 3.0, and 2.5 to 2.6 times higher with the hCBCT system, respectively. Hounsfield unit accuracy, noise, and uniformity of hCBCT exceeded the performance of the mCBCT; spatial resolution was comparable. Added x-ray beam filtration and AEC capability achieved clinical image quality for intraoperative spine surgery at reduced radiation dose to the patient in comparison to a reference O-arm system without these capabilities.

摘要

脊柱手术术中三维(3D)成像期间,需要谨慎选择方案,以控制患者辐射剂量并实现临床图像质量。本文比较了脊柱手术中常用的美敦力O型臂和配备XperCT C型臂3D锥形束CT(hCBCT)的飞利浦混合手术室的辐射剂量和图像质量。移动O型臂(mCBCT)针对四种不同患者体型(小、中、大、特大)提供三种不同的辐射剂量设置(低、标准、高)。在整个3D扫描过程中,患者的辐射剂量率保持恒定。相比之下,C-CBCT脊柱成像使用三种不同的视野(27、37和48厘米),采用自动曝光控制(AEC),根据患者厚度变化在3D扫描期间调节患者的辐射剂量率。hCBCT使用额外的X射线束过滤。对设计用于模拟脊柱和软组织的小、中、大躯干模型进行成像,以评估两种系统的辐射剂量和图像质量。考虑所有剂量设置,mCBCT对小、中、大模型成像的估计测量“患者”剂量分别为9.4-27.6毫戈瑞、8.9-33.3毫戈瑞和13.8-40.6毫戈瑞。hCBCT对相同模型成像的“患者”剂量值为2.8-4.6毫戈瑞、5.7-10.0毫戈瑞和11.0-15.2毫戈瑞。hCBCT系统对小、中及大模型的对比噪声比(CNR)分别高出2.9至3.7倍、2.0至3.0倍和2.5至2.6倍。hCBCT的亨氏单位准确性、噪声和均匀性超过了mCBCT的性能;空间分辨率相当。与没有这些功能的参考O型臂系统相比,增加的X射线束过滤和AEC功能在降低患者辐射剂量的情况下实现了术中脊柱手术的临床图像质量。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/36b3/6370984/bdfe61e8d4b8/ACM2-20-136-g001.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验