Sensors (Basel). 2024 Oct 26;24(21):6883. doi: 10.3390/s24216883.
An image-quality CT phantom was scanned with three different 3D X-ray imaging guidance devices in the operating theatre: O-Arm, Loop-X, and Airo TruCT. Default acquisition and reconstruction parameters for lumbar spine procedures were used on each device. The tube current was set to a dose level of around 27 mGy. A task-based image quality assessment was performed by calculating the noise power spectrum (NPS) and task transfer function (TTF). A detectability index (d') was calculated for three simulated bone lesions. The noise magnitude of the O-Arm was higher than the Airo TruCT, and the Loop-X had higher noise than the Airo TruCT. The highest average NPS frequency was for the O-Arm images, and the lowest was for the Loop-X. The TTFs at 50% values were similar for the Airo TruCT and Loop-X devices. Compared to Airo TruCT, the TTF at 50% value increased with the O-Arm by 53.12% and 41.20% for the Teflon and Delrin inserts, respectively. Compared to Airo TruCT, the d' value was lower with Loop-X by -26.73%, -27.02%, and -23.95% for lytic lesions, sclerotic lesions, and high-density bone, respectively. Each 3D-imaging spine surgery guidance device has its own strengths and weaknesses in terms of image quality. Cone-beam CT systems apparently offer the best compromise between noise and spatial resolution for spine surgery.
在手术室中,使用三种不同的 3D X 射线成像引导设备对一个图像质量 CT 体模进行了扫描:O-Arm、Loop-X 和 Airo TruCT。每种设备都使用了腰椎程序的默认采集和重建参数。管电流设置为约 27 mGy 的剂量水平。通过计算噪声功率谱(NPS)和任务传递函数(TTF)进行了基于任务的图像质量评估。为三个模拟骨病变计算了检测指数(d')。O-Arm 的噪声幅度高于 Airo TruCT,而 Loop-X 的噪声高于 Airo TruCT。O-Arm 图像的平均 NPS 频率最高,Loop-X 最低。Airo TruCT 和 Loop-X 设备的 TTF 在 50%值处相似。与 Airo TruCT 相比,O-Arm 的 TTF 在 50%值处增加了 53.12%和 41.20%,用于 Teflon 和 Delrin 插件。与 Airo TruCT 相比,Loop-X 的 d'值分别降低了 -26.73%、-27.02%和 -23.95%,用于溶骨性病变、硬化性病变和高密度骨。每种用于脊柱手术的 3D 成像引导设备在图像质量方面都有其自身的优势和劣势。锥形束 CT 系统显然在脊柱手术的噪声和空间分辨率之间提供了最佳折衷。
Diagn Interv Imaging. 2022-11
Injury. 2022-11
Diagnostics (Basel). 2021-8-9
Ann Transl Med. 2021-1