Abstract

PURPOSE

We evaluated the feasibility of a smartphone application-based dark adaptation (DA) measurement method (MOBILE-DA).

METHODS

On a Samsung Galaxy S8 smartphone, MOBILE-DA presented a 1.5° flashing stimulus (wavelength = 453 nm) between -1.15 and -4.33 log candela (cd)/m at 8° eccentricity using an adaptive staircase, and logged timing of user response (tapping on the screen) whenever the stimulus became visible (monocularly). In a dark room, the smartphone was placed ≈40 cm from the subject, and a white smartphone screen at maximum brightness (≈300 cd/m) for 120 seconds was used for bleaching before testing. MOBILE-DA was evaluated in normally-sighted (NV) subjects ( = 15; age, 22-82 years). Additionally, a subject with myopic retinal degeneration (MRD; VA, 20/100; age, 62 years) and another with optic nerve atrophy (ONA; visual acuity [VA], 20/500; age, 40 years) were measured. Maximum test timing was capped at 20 minutes. Linear regression was performed to determine age-effect on DA parameters: rod-cone break time ( ) and test-time ( ). Use of the normalized area under the DA characteristics (AUC) as an outcome measure was explored.

RESULTS

For NV, the repeatability coefficients for , , and AUC were ±2.1 minutes, ±5.4 minutes, and 4.4%, respectively, and aging-related delays were observed ( , = 0.47, = 0.003; , = 0.34, = 0.013; AUC, = 0.41, = 0.006). Compared to ONA and NV, DA was greatly prolonged in the MRD subject (52% larger AUC than the NV mean).

CONCLUSION

The age-effect was verified for MOBILE-DA measurements in NV subjects; impaired DA in a case with retinal-degeneration was observed.

TRANSLATIONAL RELEVANCE

This study establishes feasibility of the smartphone-based DA measurement method as a potential accessible screening tool for various vision disorders.

摘要

目的

我们评估了基于智能手机应用程序的暗适应(DA)测量方法(MOBILE - DA)的可行性。

方法

在三星Galaxy S8智能手机上,MOBILE - DA使用自适应阶梯法在8°偏心率下呈现1.5°的闪烁刺激(波长 = 453 nm),亮度范围在 - 1.15至 - 4.33对数坎德拉(cd)/m²之间,并且每当刺激可见(单眼)时记录用户响应(点击屏幕)的时间。在暗室中,将智能手机放置在距离受试者约40 cm处,在测试前使用最大亮度(约300 cd/m²)的白色智能手机屏幕进行120秒的漂白处理。MOBILE - DA在正常视力(NV)受试者(n = 15;年龄,22 - 82岁)中进行评估。此外,还测量了一名患有近视性视网膜变性(MRD;视力,20/100;年龄,62岁)的受试者和另一名患有视神经萎缩(ONA;视力[VA],20/500;年龄,40岁)的受试者。最大测试时间限制为20分钟。进行线性回归以确定年龄对DA参数的影响:视锥 - 视杆细胞转折时间( )和测试时间( )。探讨使用DA特征下的归一化面积(AUC)作为结果指标。

结果

对于NV受试者, 、 和AUC的重复性系数分别为±2.1分钟、±5.4分钟和4.4%,并且观察到与年龄相关的延迟( ,r = 0.47,P = 0.003; ,r = 0.34,P = 0.013;AUC,r = 0.41,P = 0.006)。与ONA和NV受试者相比,MRD受试者的DA大大延长(AUC比NV平均值大52%)。

结论

在NV受试者中验证了MOBILE - DA测量的年龄效应;观察到一例视网膜变性患者的DA受损。

转化相关性

本研究确立了基于智能手机的DA测量方法作为各种视力障碍潜在可及筛查工具的可行性。