Repeatability of Scotopic Sensitivity and Dark Adaptation Using a Medmont Dark-Adapted Chromatic Perimeter in Age-related Macular Degeneration.

作者信息

Uddin Durin, Jeffrey Brett G, Flynn Oliver, Wong Wai, Wiley Henry, Keenan Tiarnan, Chew Emily, Cukras Catherine

机构信息

National Eye Institute, NIH, Bethesda, Maryland, USA.

出版信息

Transl Vis Sci Technol. 2020 Jun 25;9(7):31. doi: 10.1167/tvst.9.7.31. eCollection 2020 Jun.

Abstract

PURPOSE

Functional studies of rods in age-related macular degeneration using the Medmont Dark-Adapted Chromatic Perimeter (DACP) have demonstrated impairments in scotopic sensitivities and dark adaptation (DA). We investigated the intersession repeatability of scotopic sensitivity and DA parameters including the rod intercept time recorded from the Medmont DACP.

METHODS

Scotopic thresholds (14 test points) and DA using a 30% photobleach (eight test points) were measured on two separate days from participants 50 years of age or older with a range of age-related macular degeneration severity at loci superior and inferior to the fovea. Repeatability coefficients were calculated for prebleach scotopic sensitivity, and for DA parameters including rod intercept time.

RESULTS

Twelve participants (mean age, 79.7 ± 8.1 years) repeated Medmont DACP testing within 50 days. Repeatability coefficients for prebleach scotopic sensitivity to long wavelength (red, 625 nm) and short wavelength (cyan, 505 nm) were 5.9 dB and 7.2 dB, respectively. The DA curve-derived repeatability coefficients for cone threshold was 3.9 dB, final threshold 5.3 dB, with an R value of 0.075 decades/min, rod intercept time 7.6 minutes, and RITslope 0.54 min/degree.

CONCLUSIONS

This study establishes repeatability coefficients for scotopic thresholds and multiple DA parameters obtained with the Medmont DACP in patients with age-related macular degeneration. These repeatability coefficients will serve as the basis for determining clinically meaningful change in rod function in future clinical trials.

TRANSLATIONAL RELEVANCE

Measures of repeatability parameters of scotopic thresholds and DA are essential to the accurate interpretation of results in future studies and trials using these measures.

摘要

目的

使用Medmont暗适应彩色视野计(DACP)对年龄相关性黄斑变性患者的视杆细胞进行功能研究,结果显示其暗视敏感度和暗适应(DA)存在损害。我们研究了暗视敏感度和DA参数(包括从Medmont DACP记录的视杆细胞截距时间)在不同测量时段之间的可重复性。

方法

对50岁及以上、年龄相关性黄斑变性严重程度各异的参与者,在黄斑中心凹上下部位的位点上,于两个不同日期测量暗视阈值(14个测试点)和使用30%漂白闪光的DA(8个测试点)。计算漂白前暗视敏感度以及包括视杆细胞截距时间在内的DA参数的可重复性系数。

结果

12名参与者(平均年龄79.7±8.1岁)在50天内重复进行了Medmont DACP测试。漂白前对长波长(红色,625 nm)和短波长(青色,505 nm)暗视敏感度的可重复性系数分别为5.9 dB和7.2 dB。DA曲线得出的锥体细胞阈值可重复性系数为3.9 dB,最终阈值为5.3 dB,R值为0.075十年/分钟,视杆细胞截距时间为7.6分钟,RIT斜率为0.54分钟/度。

结论

本研究确定了年龄相关性黄斑变性患者使用Medmont DACP获得的暗视阈值和多个DA参数的可重复性系数。这些可重复性系数将作为未来临床试验中确定视杆细胞功能具有临床意义变化的基础。

转化相关性

暗视阈值和DA可重复性参数的测量对于准确解释未来使用这些测量方法的研究和试验结果至关重要。