两种不同吸氧水平下呼吸测热仪的可靠性和一致性研究。

Reliability of, and Agreement Between, two Breath-by-Breath Indirect Calorimeters at Varying Levels of Inspiratory Oxygen.

机构信息

Respiratory and Critical Care Group, Department of Health Science and Technology, Aalborg University, Aalborg, Denmark.

Department of Anesthesiology, Aalborg University Hospital, Aalborg, Denmark.

出版信息

Nutr Clin Pract. 2019 Oct;34(5):767-774. doi: 10.1002/ncp.10250. Epub 2019 Feb 4.

Abstract

BACKGROUND

Indirect calorimetry (IC) is considered the accurate way of measuring energy expenditure (EE). IC devices often apply the Haldane transformation, introducing errors at inspiratory oxygen fraction (FiO ) >60%. The aim was to assess measurement reliability and agreement between an unevaluated IC (device 2) (Beacon Caresystem, Mermaid Care A/S, Noerresundby, Denmark) not using Haldane transformation and an IC that does (device 1) (Ecovx, GE, Helsinki, Finland) at varying FiO .

METHODS

Twenty healthy male subjects participated, with 16 completing the study (33 ± 9 years, 83.3 ± 16 kg, 1.83 ± 0.08 m). Subjects were mechanically ventilated in pressure support (3cmH O; positive end-expiratory pressure: 3cmH O) at FiO of 21%, 50%, 85%, and 21% for 15 minutes at each FiO . Mean EE, oxygen consumption (VO ), and CO production (VCO ) were compared within and between devices across FiO levels.

RESULTS

Device 2 showed within-device EE significant differences at 21% vs 50% FiO and device 1 for VCO at 50% vs. 85% FiO . For all variables, both devices showed reliable measurements at 21% and 50% FiO , but at 85%, FiO bias and limits of agreement increased. Between devices, there were significant differences for EE at both 21% and 85% FiO for VO and for VCO at 85% FiO .

CONCLUSION

Both systems measured EE, VO , and VCO at 21%-85% FiO reliably but with bias at 85% FiO . The devices were in agreement at 21% and 50% FiO , but further studies need to confirm accuracy at high FiO .

摘要

背景

间接热量测定法(IC)被认为是测量能量消耗(EE)的准确方法。IC 设备通常应用 Haldane 变换,在吸入氧分数(FiO )>60%时会引入误差。目的是评估一种未经评估的 IC(设备 2)(Beacon Caresystem,Mermaid Care A/S,Noerresundby,丹麦)与一种使用 Haldane 变换的 IC(设备 1)(Ecovx,GE,Helsinki,芬兰)在不同 FiO 下的测量可靠性和一致性。

方法

20 名健康男性受试者参与,16 名完成研究(33±9 岁,83.3±16kg,1.83±0.08m)。受试者在压力支持(3cmH 2 O;呼气末正压:3cmH 2 O)下机械通气,FiO 分别为 21%、50%、85%和 21%,每种 FiO 下通气 15 分钟。在 FiO 水平内和设备间比较平均 EE、耗氧量(VO )和 CO 生成量(VCO )。

结果

设备 2 在 FiO 为 21%与 50%时显示 EE 有显著差异,设备 1 在 FiO 为 50%与 85%时显示 VCO 有显著差异。对于所有变量,两个设备在 FiO 为 21%和 50%时均显示可靠的测量结果,但在 FiO 为 85%时,FiO 偏差和一致性界限增加。两个设备之间,在 FiO 为 21%和 85%时,VO 的 EE 以及 FiO 为 85%时的 VCO 有显著差异。

结论

两个系统均能可靠地测量 FiO 为 21%-85%时的 EE、VO 和 VCO,但 FiO 为 85%时存在偏差。两个设备在 FiO 为 21%和 50%时一致,但需要进一步研究证实高 FiO 时的准确性。