使用改良弯曲试验对自锁托槽中涂层和未涂层镍钛丝的载荷-挠度特性进行的一项研究。

Load-deflection characteristics of coated and noncoated nickel-titanium wires in self-ligating brackets using a modified bending test: An study.

作者信息

Tikku Tripti, Khanna Rohit, Agarwal Akhil, Srivastava Kamna, Shekhar Shashank, Shukla Ivy

机构信息

Department of Orthodontics, Babu Banarasi Das College of Dental Sciences, Lucknow, Uttar Pradesh, India.

Department of Material Sciences, IIT Kanpur, Kanpur, Uttar Pradesh, India.

出版信息

Dent Res J (Isfahan). 2019 Jan-Feb;16(1):1-6.

Abstract

BACKGROUND

To determine and compare the force-deflection values of different types of nickel-titanium (NiTi) wires during unloading phase at varying deflections, that is 1 mm, 1.5 mm, 2 mm, and 2.5 mm, with the use of self-ligating ceramic brackets using modified bending test on a typodont under controlled temperature conditions.

MATERIALS AND METHODS

In this study total of 45 wires of ovoid shape of three different NiTi wires - pseudoelastic NiTi (Group I), heat activated NiTi (Group II), and esthetic coated NiTi (Group III) for maxillary arch were tested after insertion in ceramic self-ligating brackets bonded to plastic teeth of phantom jaw. The maxillary left lateral incisor was removed to simulate a malpositioned tooth which acted as the load site, and load-deflection characteristics were measured during unloading using Instron, and data analyzed statically by two-way analysis of variance, Tukey's test, intraclass correlation coefficient and Pearson correlation coefficient. A two-sided (α = 2) < 0.05 was considered statistically significant.

RESULTS

When wires were compared at each deflection statistically significant difference was observed between the three groups of wires (Group I > Group II > Group III) at all the four levels of deflection except for Group II versus Group III at 1 mm, 1.5 mm, and 2 mm of deflection. Statistically significant difference was noted in mean load values for comparisons made at different deflections for each wire except for the comparison made at 1.5 versus 2 mm for Group II and Group III.

CONCLUSION

Overall comparison showed esthetic coated Ni-Ti wires gave significantly lower mean load values, followed by heat activated and pseudoelastic NiTi wires. Thus, heat activated NiTi wires are best suited in patients with severe malpositions/periodontitis, while for esthetically conscious patients esthetic coated NiTi can be used.

摘要

背景

在可控温度条件下,使用改良弯曲试验,在模型牙上对使用自锁陶瓷托槽的不同类型镍钛(NiTi)丝在卸载阶段(即1毫米、1.5毫米、2毫米和2.5毫米的不同挠度)的力-挠度值进行测定和比较。

材料与方法

本研究中,将三种不同的上颌卵圆形NiTi丝——伪弹性NiTi(第一组)、热激活NiTi(第二组)和美观涂层NiTi(第三组)共45根丝插入粘结到模拟颌骨塑料牙上的陶瓷自锁托槽后进行测试。拔除上颌左侧侧切牙以模拟错位牙作为加载部位,使用英斯特朗万能材料试验机在卸载过程中测量载荷-挠度特性,并通过双向方差分析、Tukey检验、组内相关系数和Pearson相关系数进行统计学分析。双侧(α = 2)<0.05被认为具有统计学显著性。

结果

在每个挠度下对丝进行比较时,除了在1毫米、1.5毫米和2毫米挠度时第二组与第三组之间外,在所有四个挠度水平下,三组丝之间均观察到统计学显著性差异(第一组>第二组>第三组)。除了第二组和第三组在1.5毫米与2毫米挠度之间的比较外,对于每种丝在不同挠度下进行比较时,平均载荷值存在统计学显著性差异。

结论

总体比较表明,美观涂层Ni-Ti丝的平均载荷值显著较低,其次是热激活NiTi丝和伪弹性NiTi丝。因此,热激活NiTi丝最适合严重错位/牙周炎患者,而对于注重美观的患者,可以使用美观涂层NiTi丝。