Department of Urology, Institute of Urology, Beijing Chaoyang Hospital, Capital Medical University, Beijing, 100020, People's Republic of China.
Department of Urology, Beijing Hospital, National Center of Gerontology, Beijing, 100730, People's Republic of China.
Int Urol Nephrol. 2019 Apr;51(4):627-632. doi: 10.1007/s11255-019-02115-3. Epub 2019 Feb 27.
To explore the feasibility and safety of the Tsinghua PINS Remote Tech to facilitate sacral neuromodulation programming procedure.
For 22 patients who had previously participated in the phase III clinical trial for treating overactive bladder with the Tsinghua PINS sacral neuromodulation system during several Hospital, PINS Remote Tech was applied to perform postoperative parameter adjustment in order to evaluate the safety and reliability of this new technique. Telephone surveys on Remote Tech-related questionnaires were also conducted.
17/22 patients underwent 26 parameter adjustments, average adjustment frequency was 1.53 times per person; the average adjustment time was 23.4 ± 5.1 min (15-32 min). The total effective rate of the Remote control was 14/17 (82.3%). 7/17 (41.1%) patients' symptoms recurrence due to not knowing how to handle patient controller, these patients were instructed on how to use it correctly through Remote Tech even without reprogramming it. Other 10 patients received reprogramming. There was no discomfort during and after parameter adjustment. The questionnaire survey showed that the remote technology saved patients' time and lowered financial costs, significantly improved patient satisfaction. All patients expressed their willingness to recommend it to other patients.
The PINS Remote Tech can significantly reduce the financial cost and provide a remote reprogram control service that is as safe and reliable as outpatient program control.
探索清华 PINS Remote Tech 辅助骶神经调控程控的可行性和安全性。
对 22 例曾参与清华 PINS 骶神经调控系统治疗膀胱过度活动症 III 期临床试验的患者,应用 PINS Remote Tech 进行术后参数调整,以评估该新技术的安全性和可靠性。同时进行与远程技术相关的问卷调查。
22 例患者中 17 例进行了 26 次参数调整,平均调整频率为 1.53 次/人;平均调整时间为 23.4±5.1min(15-32min)。远程控制的总有效率为 14/17(82.3%)。由于不知道如何处理患者控制器,7/17(41.1%)患者症状复发,通过远程技术指导他们正确使用,即使不重新编程也能解决问题。其他 10 名患者接受了重新编程。参数调整过程中和调整后均无不适。问卷调查显示,远程技术节省了患者的时间和经济成本,显著提高了患者的满意度。所有患者均表示愿意向其他患者推荐。
PINS Remote Tech 可显著降低经济成本,并提供安全可靠的远程程控服务。