使用Corvis ST测量眼压并与压平眼压计和Tomey非接触眼压计进行比较。

Intraocular pressure measurement with Corvis ST in comparison with applanation tonometry and Tomey non-contact tonometry.

作者信息

Luebke Jan, Bryniok L, Neuburger M, Jordan J F, Boehringer D, Reinhard T, Wecker T, Anton A

机构信息

Eye Center, Medical Center, University Hospital Freiburg, University of Freiburg, Killianstrasse 5, 79106, Freiburg, Germany.

Faculty of Medicine, Albert Ludwigs University of Freiburg, Freiburg, Germany.

出版信息

Int Ophthalmol. 2019 Nov;39(11):2517-2521. doi: 10.1007/s10792-019-01098-5. Epub 2019 Apr 9.

Abstract

PURPOSE

Intraocular pressure (IOP) measurement can be performed with different methods. Newer methods have to be compared to the standard method, the Goldmann applanation tonometry (GAT). We herein compare two air-puff tonometers, the non-contact tonometer (Tomey NCT) and the Corvis ST (CST) with GAT in eyes with a broad spectrum of IOP.

METHODS

Two hundred and forty-nine eyes of 249 patients (with diagnosis of either glaucoma or ocular hypertension) were included in this monocenter prospective cohort study. Each eye underwent IOP measurements via GAT, NCT and CST. Bland-Altman plots were calculated to compare the different methods in the three groups. Paired t tests were used for statistical comparison between the three measurement methods. The difference between the different methods was tested on correlation against central corneal thickness (CCT).

RESULTS

Mean IOP in GAT was 17.6 mmHg (standard deviation (SD) 5.9), 16.3 mmHg (SD 5.6) in NCT and 18.0 mmHg (SD 5.5) in CST. Comparisons between GAT and CST vs. NCT showed significant differences (p < 0.001), while GAT vs. CST showed no significant difference (p = 0.1162). Mean CCT was 538.7 µm (SD 35.1).

CONCLUSIONS

Mean values of GAT and CST show comparable results. However, both GAT and CST differ significantly from NCT. NCT shows lower IOP values compared to both other methods.

摘要

目的

眼压(IOP)测量可通过不同方法进行。新方法必须与标准方法——Goldmann压平眼压计(GAT)进行比较。我们在此比较两种气流眼压计,即非接触眼压计(Tomey NCT)和Corvis ST(CST)与GAT在眼压范围广泛的眼中的情况。

方法

本单中心前瞻性队列研究纳入了249例患者的249只眼(诊断为青光眼或高眼压症)。每只眼均通过GAT、NCT和CST进行眼压测量。计算Bland-Altman图以比较三组中的不同方法。配对t检验用于三种测量方法之间的统计学比较。不同方法之间的差异针对中央角膜厚度(CCT)进行相关性检验。

结果

GAT测量的平均眼压为17.6 mmHg(标准差(SD)5.9),NCT为16.3 mmHg(SD 5.6),CST为18.0 mmHg(SD 5.5)。GAT和CST与NCT之间的比较显示出显著差异(p < 0.001),而GAT与CST之间无显著差异(p = 0.1162)。平均CCT为538.7 µm(SD 35.1)。

结论

GAT和CST的平均值显示出可比的结果。然而,GAT和CST与NCT均有显著差异。与其他两种方法相比,NCT显示出较低的眼压值。