Phys Med Biol. 2019 May 29;64(11):115014. doi: 10.1088/1361-6560/ab1856.
The RAYCAN Trans-PET/CT X5 is a preclinical positron emission tomography and computed tomography (PET/CT) system intended for in vivo imaging of rats and mice, featuring all-digital readout electronics for PET data acquisition. The National Electrical Manufacturers Association (NEMA) NU 4-2008 performance evaluation was conducted on the RAYCAN Trans-PET/CT X5 in addition to assessing in vivo imaging performance of the system on live animals. The performance characteristics of the system were evaluated, including system spatial resolution, count rate performance, sensitivity and image quality. The system imaging performance is assessed in dynamic in vivo PET imaging. The system resolution defined as full width half maximum (FWHM) was 2.07 mm, 2.11 mm and 1.31 mm for the tangential, radial and axial resolution, respectively, at the center of the field of view. The peak noise equivalent count rate (NECR) values measured were 61 kcps at 0.19 MBq ml for the rat size phantom and 126 kcps at 1.53 MBq ml for the mouse size phantom. Scatter fractions were 24% and 14% for the rat and mouse phantom. The measured peak sensitivity of the system was 1.70%. Image quality in static imaging was deemed sufficient based on the image quality phantom study, with average activity concentration of 155 ± 8.6 kBq ml and image uniformity of 5.57% when using two-dimensional filtered backprojection algorithm (2D-FBP). Rods in the image quality phantom were visualized easily up to 2 mm in size. In dynamic in vivo PET imaging, time-activity-curves from several regions were successfully measured, characterizing the radioactivity distribution in myocardial blood pool, liver, left ventricle and the lung. In conclusion, the RAYCAN Trans-PET/CT X5 system can be considered a suitable option for basic imaging needs in preclinical imaging.
RAYCAN Trans-PET/CT X5 是一款用于大鼠和小鼠体内成像的临床前正电子发射断层扫描和计算机断层扫描(PET/CT)系统,采用全数字读出电子技术进行 PET 数据采集。除了评估系统在活体动物上的体内成像性能外,还按照美国电器制造商协会(NEMA)NU 4-2008 性能评估标准对 RAYCAN Trans-PET/CT X5 进行了评估。对系统的性能特征进行了评估,包括系统空间分辨率、计数率性能、灵敏度和图像质量。在动态体内 PET 成像中评估了系统的成像性能。系统分辨率定义为半峰全宽(FWHM),在视场中心处,切向、径向和轴向分辨率分别为 2.07、2.11 和 1.31mm。在大鼠尺寸体模中测量到的最高噪声等效计数率(NECR)值为 61 kcps,在鼠尺寸体模中为 126 kcps。在大鼠和小鼠体模中,散射分数分别为 24%和 14%。系统的测量峰值灵敏度为 1.70%。基于图像质量体模研究,静态成像中的图像质量被认为是足够的,使用二维滤波反投影算法(2D-FBP)时,平均活度浓度为 155±8.6 kBq ml,图像均匀性为 5.57%。图像质量体模中的棒很容易被可视化,大小可达 2mm。在动态体内 PET 成像中,成功测量了来自几个区域的时间-活度曲线,描绘了心肌血池、肝脏、左心室和肺部的放射性分布。总之,RAYCAN Trans-PET/CT X5 系统可被视为临床前成像中基本成像需求的合适选择。