Digital Biograph Vision PET/CT 系统的性能特点。

Performance Characteristics of the Digital Biograph Vision PET/CT System.

机构信息

Department of Nuclear Medicine and Molecular Imaging, University Medical Center Groningen, University of Groningen, Groningen, The Netherlands

Department of Nuclear Medicine and Molecular Imaging, University Medical Center Groningen, University of Groningen, Groningen, The Netherlands.

出版信息

J Nucl Med. 2019 Jul;60(7):1031-1036. doi: 10.2967/jnumed.118.215418. Epub 2019 Jan 10.

Abstract

This study evaluated the performance of the Biograph Vision digital PET/CT system according to the NEMA NU 2-2012 standard (published by the National Electrical Manufacturers Association [NEMA]) to allow for a reliable, reproducible, and intersystem-comparable performance measurement. The new digital PET/CT system features silicon photomultiplier-based detectors with 3.2-mm lutetium oxyorthosilicate crystals and full coverage of the scintillator area. The PET components incorporate 8 rings of 38 detector blocks, and each block contains 4 × 2 mini blocks. Each mini block consists of a 5 × 5 lutetium oxyorthosilicate array of 3.2 × 3.2 × 20 mm crystals coupled to a silicon photomultiplier array of 16 × 16 mm, resulting in an axial field of view of 26.1 cm. In this study, PET/CT system performance was evaluated for conformation with the NEMA NU 2-2012 standard, with additional measurements described in the new NEMA NU 2-2018 standard. Spatial resolution, sensitivity, count-rate performance, accuracy of attenuation and scatter correction, image quality, coregistration accuracy, and time-of-flight performance were determined. Measurements were directly compared with results from its predecessor, the Biograph mCT Flow, using existing literature. Moreover, feasibility to comply with the European Association of Nuclear Medicine Research Ltd. (EARL) criteria was evaluated, and some illustrative patient PET images were obtained. The Biograph Vision showed a transverse and axial spatial resolution of 3.6 and 3.5 mm, respectively, in full width at half maximum at a 1-cm offset from the center of the field of view (measured with a Na 0.25-mm point source), a NEMA sensitivity of 16.4 kcps/MBq, and a NEMA peak noise-equivalent count-rate of 306 kcps at 32 kBq/mL. Time-of-flight resolution varied from 210 to 215 as count-rate increased up to the peak noise-equivalent count-rate. The overall image contrast seen with the NEMA image quality phantom ranged from 77.2% to 89.8%. Furthermore, the system was able to comply with the current and future EARL performance criteria. The Biograph Vision outperforms the analog Biograph mCT Flow, and the system is able to meet European harmonizing performance standards.

摘要

本研究根据美国电器制造商协会(NEMA)发布的 NEMA NU 2-2012 标准(以下简称“标准”)评估了 Biograph Vision 数字 PET/CT 系统的性能,以实现可靠、可重复且系统间可比的性能测量。新型数字 PET/CT 系统采用基于硅光电倍增器的探测器,具有 3.2mm 硅酸镥(LSO)晶体和整个闪烁体区域全覆盖。PET 组件包含 8 个 38 个探测器模块的环,每个模块包含 4×2 个迷你模块。每个迷你模块由一个 5×5 的 LSO 阵列组成,该阵列由 3.2×3.2×20mm 的晶体组成,并与一个 16×16mm 的硅光电倍增器阵列耦合,轴向视野为 26.1cm。在这项研究中,根据标准评估了 PET/CT 系统的性能,并根据新标准 NEMA NU 2-2018 进行了额外的测量。确定了空间分辨率、灵敏度、计数率性能、衰减和散射校正准确性、图像质量、配准精度和飞行时间性能。使用现有文献,直接将测量结果与前一代产品 Biograph mCT Flow 的结果进行了比较。此外,评估了符合欧洲核医学研究协会(EARL)标准的可行性,并获得了一些示踪 PET 图像。Biograph Vision 的中心和边缘的横向和轴向空间分辨率分别为 3.6mm 和 3.5mm(在视野中心 1cm 处用 Na 0.25mm 点源测量),NEMA 灵敏度为 16.4kcps/MBq,在 32kBq/mL 时 NEMA 峰噪声等效计数率为 306kcps。随着计数率增加到峰值噪声等效计数率,飞行时间分辨率从 210 到 215 变化。使用 NEMA 图像质量体模观察到的整体图像对比度在 77.2%到 89.8%之间。此外,该系统能够符合当前和未来的 EARL 性能标准。Biograph Vision 优于模拟 Biograph mCT Flow,并且该系统能够满足欧洲协调性能标准。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验