Sensors (Basel). 2019 Apr 13;19(8):1781. doi: 10.3390/s19081781.
(1) Background: Measuring joint range of motion has traditionally occurred with a universal goniometer or expensive laboratory based kinematic analysis systems. Technological advances in wearable inertial measurement units (IMU) enables limb motion to be measured with a small portable electronic device. This paper aims to validate an IMU, the 'Biokin', for measuring shoulder range of motion in healthy adults; (2) Methods: Thirty participants completed four shoulder movements (forward flexion, abduction, and internal and external rotation) on each shoulder. Each movement was assessed with a goniometer and the IMU by two testers independently. The extent of agreement between each tester's goniometer and IMU measurements was assessed with intra-class correlation coefficients (ICC) and Bland-Altman 95% limits of agreement (LOA). Secondary analysis compared agreement between tester's goniometer or IMU measurements (inter-rater reliability) using ICC's and LOA; (3) Results: Goniometer and IMU measurements for all movements showed high levels of agreement when taken by the same tester; ICCs > 0.90 and LOAs < ±5 degrees. Inter-rater reliability was lower; ICCs ranged between 0.71 to 0.89 and LOAs were outside a prior defined acceptable LOAs (i.e., > ±5 degrees); (4) Conclusions: The current study provides preliminary evidence of the concurrent validity of the Biokin IMU for assessing shoulder movements, but only when a single tester took measurements. Further testing of the Biokin's psychometric properties is required before it can be confidently used in routine clinical practice and research settings.
(1) 背景:传统上,关节活动范围的测量是使用通用量角器或昂贵的实验室运动分析系统进行的。可穿戴惯性测量单元 (IMU) 的技术进步使我们能够使用小型便携式电子设备来测量肢体运动。本文旨在验证一种用于测量健康成年人肩部活动范围的 IMU,即“Biokin”;(2) 方法:30 名参与者在每个肩部上完成了四个肩部运动(前屈、外展、内收和外旋)。每个运动均由两个测试人员使用量角器和 IMU 分别进行评估。通过使用组内相关系数 (ICC) 和 Bland-Altman 95%一致性界限 (LOA) 评估了每位测试员的量角器和 IMU 测量值之间的一致性程度;(3) 结果:当由同一位测试员进行测量时,所有运动的量角器和 IMU 测量值均显示出高度的一致性;ICC>0.90,LOA<±5 度。测试员之间的可靠性较低;ICC 范围在 0.71 到 0.89 之间,LOA 在之前定义的可接受 LOA 之外(即,>±5 度);(4) 结论:本研究初步证明了 Biokin IMU 在评估肩部运动时的同时效度,但前提是由单个测试员进行测量。在可以自信地将 Biokin 用于常规临床实践和研究环境之前,需要进一步测试其心理测量特性。