In-situ breakage of an implantable venous access system.

作者信息

Saifi J, MacDowell R T, Khouri G, Webster D J

出版信息

Eur J Surg Oncol. 1987 Apr;13(2):159-62.

PMID:3104094
Abstract

A totally implantable system for venous access (Port-A-Cath, Pharmacia, Nu-tech) has now entered clinical use. This system consists of a silastic catheter secured to a stainless steel reservoir with a locking ring. The system has been shown to have several advantages when compared with earlier venous access systems. However, the advantages of total implantation have also disposed the system to new complications. Two cases are presented of patients who had this system placed and developed in-situ separation of the catheter from the reservoir. The catheter was removed surgically from the subcutaneous tissues in one patient and with a Dormain basket in the second because complete migration of the catheter into the central venous system had occurred. In neither patient could the precise cause for the separation be ascertained. However, it can be appreciated that if chemotherapeutic agents were infused in these patients the results would have been disastrous. It is recommended that if blood cannot be easily aspirated after gentle to and fro flushes with a 10 ml syringe, a chest radiograph be obtained prior to the infusion of any substance to check for continuity in the system. Any breaks in the system should be immediately repaired since migration of the catheter into the venous system can occur.

摘要

一种完全植入式静脉通路系统(Port - A - Cath,法玛西亚公司,Nu - tech)现已投入临床使用。该系统由一根硅橡胶导管和一个带锁环的不锈钢贮液器组成,导管固定在贮液器上。与早期的静脉通路系统相比,该系统已显示出若干优势。然而,完全植入的优势也使该系统出现了一些新的并发症。本文介绍了两例植入该系统后发生导管与贮液器原位分离的患者。其中一例患者通过手术从皮下组织取出导管,另一例则使用Dormain网篮取出,因为导管已完全移入中心静脉系统。两名患者均无法确定分离的确切原因。然而,可以想象,如果在这些患者中输注化疗药物,后果将不堪设想。建议在使用10毫升注射器轻轻来回冲洗后,如果不易抽出血液,在输注任何物质之前应进行胸部X光检查,以检查系统的连续性。由于导管可能会移入静脉系统,系统中的任何破损都应立即修复。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验