Suppr超能文献

坦桑尼亚的质量控制措施:实施了吗?

Quality Control Measures in Tanzania: Is it Done?

作者信息

Ngoye Wilson M, Motto Jenny A, Muhogora Wilbroad E

机构信息

Department of Radiography, University of Johannesburg, Johannesburg, South Africa; Department of Ionizing Radiation, Tanzania Atomic Energy Commission, Arusha, Tanzania.

Department of Radiography, University of Johannesburg, Johannesburg, South Africa.

出版信息

J Med Imaging Radiat Sci. 2015 Sep;46(3S):S23-S30. doi: 10.1016/j.jmir.2015.06.004.

Abstract

BACKGROUND

Regular implementation of a quality control (QC) programme in diagnostic radiology is vital if consistent optimal equipment performance, quality images, and accurate diagnosis at optimum radiation dose and costs are to be achieved. This highlighted the necessity for the Tanzania Atomic Energy Commission (TAEC) to establish a training programme to enable radiographers to implement a QC programme in their departments because there are no clinical medical physicists in diagnostic radiology in the country. However, the status of programme implementation is not precisely known.

AIM

The aim of this study was to investigate the status of implementation of the QC measures as performed by radiographers in diagnostic radiography departments in Tanzania.

METHOD

A quantitative cross-sectional survey, using a questionnaire, was conducted on a sample of 84 radiographers who had been trained in the QC programme and who were practising in 54 hospitals within Tanzania.

RESULTS

A large number of respondents were not implementing the following QC tests: tube output, kV, mAs and timer (94%), collimation (53.5%), and densitometry and sensitometry (87.7%). The tests for film viewing box and lead rubber protective apparel were not implemented by 64.2% and 59% of the respondents, respectively. The cassette inspection and darkroom inspection were reported as being implemented by most respondents, although the testing was not observing the recommended schedule. Furthermore, the departments had no records and procedures for the QC programme, and only the locally improvized QC test tools were reported to be available.

CONCLUSION

The QC training objectives of TAEC have not been largely achieved because radiographers are not adequately implementing the QC programme. Without the QC programme in place, equipment malfunctioning will not be revealed at an early stage, and this can cause poor image quality and/or higher dose to patients. Therefore, active involvement of the TAEC, the ministry of health, hospital management teams, and radiographers is desirable to achieve full implementation of the programme.

摘要

背景

如果要在最佳辐射剂量和成本下实现一致的最佳设备性能、高质量图像和准确诊断,那么在诊断放射学中定期实施质量控制(QC)计划至关重要。这凸显了坦桑尼亚原子能委员会(TAEC)建立培训计划的必要性,以使放射技师能够在其部门实施质量控制计划,因为该国诊断放射学领域没有临床医学物理学家。然而,该计划的实施状况尚不清楚。

目的

本研究的目的是调查坦桑尼亚诊断放射科放射技师实施质量控制措施的状况。

方法

采用问卷调查的定量横断面调查方法,对84名接受过质量控制计划培训并在坦桑尼亚54家医院工作的放射技师进行了抽样调查。

结果

大量受访者未实施以下质量控制测试:管输出、千伏、毫安秒和定时器(94%)、准直(53.5%)以及密度测定和感光度测定(87.7%)。分别有64.2%和59%的受访者未实施观片箱和铅橡胶防护服测试。大多数受访者报告已实施暗盒检查和暗室检查,尽管测试未遵循推荐的时间表。此外,各部门没有质量控制计划的记录和程序,仅报告有当地改进的质量控制测试工具。

结论

TAEC的质量控制培训目标在很大程度上尚未实现,因为放射技师未充分实施质量控制计划。如果没有质量控制计划,设备故障将无法在早期被发现,这可能导致图像质量差和/或给患者带来更高剂量。因此,TAEC、卫生部、医院管理团队和放射技师积极参与,以实现该计划的全面实施是可取的。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。