Suppr超能文献

坦桑尼亚根据质量控制性能的诊断X射线设备

Diagnostic X-ray facilities as per quality control performances in Tanzania.

作者信息

Sungita Yesaya Yohana

机构信息

Tanzania Atomic Energy Commission, Arusha, Tanzania.

出版信息

J Appl Clin Med Phys. 2006 Nov 28;7(4):66-73. doi: 10.1120/jacmp.v7i4.2291.

Abstract

Without appropriate quality control (QC) and preventative maintenance (PM) measures for X-ray machines in place, the benefits of reduced dose to the patient and early diagnosis will not be realized. Quality control and PM also make it possible to unify X-ray-imaging practices in the country using international image quality guidelines. The impetus for the present work resulted from the concern that with the recent increase in the numbers of X-ray machines in Tanzania, but with limited technical support to maintain and operate them, can increase radiation risk to patients and lower diagnostic accuracy. The aim of this work is to report on the current status of diagnostic X-ray machines in Tanzania in order to produce the data needed to formulate QC and PM policies and strategies. These policies and strategies are needed to ensure that patients receive the lowest possible radiation risk and maximum health benefits from X-ray examinations. Four QC tests were performed on a total of 196 X-ray units. Accurate beam alignment and collimation were tested on 80 (41%) units, the timer accuracy was tested on 120 (61%) units, and a radiation leakage test was performed on 47(24%) units. Preventative maintenance tests were performed on all 196 X-ray units. The results showed that of the units tested for QC, 59% failed the kilovoltage (kVp) test, 57% failed the timer accuracy test, 60% failed the beam alignment test, and 20% failed the radiation leakage test. Only 13% of the units passed the PM test: 53% of the units were defective, and 34% were out of order. As a result of the PM findings, the government has introduced a rehabilitation project to service X-ray units and replace nonoperational X-ray units. The new units have full support service contracts signed by their suppliers. As a result of the QC findings, X-ray maintenance retraining programs have been introduced.

摘要

如果没有针对X光机的适当质量控制(QC)和预防性维护(PM)措施,就无法实现降低患者剂量和早期诊断的益处。质量控制和预防性维护还使得利用国际图像质量指南统一该国的X光成像操作成为可能。开展本研究的动力源于这样一种担忧,即随着坦桑尼亚近期X光机数量的增加,但维护和操作这些机器的技术支持有限,可能会增加患者的辐射风险并降低诊断准确性。这项工作的目的是报告坦桑尼亚诊断用X光机的现状,以便生成制定质量控制和预防性维护政策及策略所需的数据。需要这些政策和策略来确保患者在接受X光检查时辐射风险降至最低,并获得最大的健康益处。对总共196台X光设备进行了四项质量控制测试。对80台(41%)设备测试了准确的光束对准和准直,对120台(61%)设备测试了定时器精度,对47台(24%)设备进行了辐射泄漏测试。对所有196台X光设备都进行了预防性维护测试。结果显示,在接受质量控制测试的设备中,59%的设备千伏(kVp)测试不合格,57%的设备定时器精度测试不合格,60%的设备光束对准测试不合格,20%的设备辐射泄漏测试不合格。只有13%的设备通过了预防性维护测试:53%的设备有缺陷,34%的设备无法正常运行。根据预防性维护的结果,政府启动了一个修复项目,为X光设备提供维修服务并更换无法使用的X光设备。新设备有供应商签署的全面支持服务合同。根据质量控制的结果,已经推出了X光维护再培训项目。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/d6c4/5722384/82bb56348c80/ACM2-7-066-g001.jpg

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验