Anterior Segment Optical Coherence Tomography Signs of Local Dilatation Effect of a Micro-Stent on Schlemm's Canal.

作者信息

Gillmann Kevin, Bravetti Giorgio Enrico, Mansouri Kaweh, Mermoud André

机构信息

Glaucoma Research Center, Montchoisi Clinic, Swiss Vision Network, Lausanne.

出版信息

Nepal J Ophthalmol. 2018 Jul;10(20):184-187. doi: 10.3126/nepjoph.v10i2.23030.

Abstract

INTRODUCTION

The iStent inject® (Glaukos Corporation, CA, USA) is a relatively new device designed to be implanted ab-interno through the trabecular meshwork. This is, to the best of our knowledge, the first in-vivo description of a trabecular bypass device visualised with anterior segment optical coherence tomography (AS-OCT), and report of its structural effect on Schlemm's canal.

CASE REPORT

A 74 year-old female patient suffering from long-standing primary open-angle glaucoma and nuclear sclerosis underwent cataract surgery combined with the implantation of two iStent injects®. Surgery was uncomplicated and achieved intraocular pressure (-1 mmHg) and medication (-2 molecules) reduction at 6 months. Under AS-OCT (Spectralis OCT, Heidelberg Engineering AG, Germany) the stent appears as a 300 μm long hyper reflective hollow device within the trabecular meshwork. Approximately a third of it protruded into the anterior chamber. Profound OCT signal loss was notable within the shadow of the device. A second AS-OCT section 500 μm beside the microstent shows a markedly dilated Schlemm's canal, with a major diameter of 390 μm.

DISCUSSIONS

This report confirms that AS-OCT is a suitable technique to assess microstent positioning, and provides a first report on the in-vivo appearance of a functioning stent. It also indicates that iStent injects® could have a tangible effect on adjacent portions of Schlemm's canal with, in this case, a 220% increase in canal diameter compared to the observed average (122 μm). This suggests the IOP-lowering effect of trabecular bypass devices could rely on a dual mechanism involving Schlemm's canal dilatation.

摘要

引言

iStent inject®(美国加利福尼亚州Glaukos公司)是一种相对较新的装置,设计用于通过小梁网进行内植入。据我们所知,这是首次通过眼前节光学相干断层扫描(AS-OCT)对小梁旁路装置进行体内描述,并报告其对施莱姆管的结构影响。

病例报告

一名74岁患有长期原发性开角型青光眼和核硬化的女性患者接受了白内障手术,并植入了两枚iStent inject®。手术过程顺利,6个月时眼压降低了1 mmHg,用药减少了2种。在AS-OCT(德国海德堡工程公司的Spectralis OCT)检查下,支架在小梁网内呈现为一个300μm长的高反射中空装置。大约三分之一的支架突出到前房。在装置的阴影内可见明显的OCT信号丢失。在微支架旁500μm处的另一张AS-OCT图像显示施莱姆管明显扩张,最大直径为390μm。

讨论

本报告证实AS-OCT是评估微支架定位的合适技术,并首次报告了功能性支架的体内外观。它还表明,iStent inject®可能对施莱姆管的相邻部分产生切实影响,在本病例中,与观察到的平均直径(122μm)相比,管直径增加了220%。这表明小梁旁路装置降低眼压的作用可能依赖于涉及施莱姆管扩张的双重机制。