Department of General, Oncological and Gastrointestinal Surgery, Orlowski Hospital, Centre of Postgraduate Medical Education, Warsaw, Poland.
Warsaw University of Technology, Faculty of Mechatronics, Institute of Metrology and Biomedical, Warsaw, Poland.
Sci Rep. 2019 May 8;9(1):7097. doi: 10.1038/s41598-019-43637-1.
Obstetric anal sphincter injuries are the most common cause of fecal incontinence in women yet remain under-diagnosed. The aim of this study was to assess the suitability of impedance spectroscopy for diagnosing sphincter injuries arising during delivery. This was a prospective single-center study. 22 female patients were included: 10 with symptoms of sphincter dysfunction, in the early postpartum period, and 12 unaffected, in the distant period of more than 2 years after natural delivery. The presence, extent and severity of anal sphincters injury was assessed by measuring the sphincter parameters in physical examination, the degree of sphincter damage in endoanal ultrasound imaging and the sphincters function parameters in anorectal manometry. All measurements were used as references and compared with the outcomes from the impedance spectroscopy models. Impedance spectroscopy showed the highest precision (with mean accuracy of 83.9%) in relation to transanal ultrasonography. 74.1% of its results corresponded to the results of rectal physical examination and 76.7% - to those of anorectal manometry. The method showed the highest accuracy in the assessment of the sphincter's parameters, both anatomically and functionally. New impedance spectroscopy techniques hold promise for detecting obstetric anal sphincter injuries.
产科肛门括约肌损伤是女性粪便失禁最常见的原因,但仍未得到充分诊断。本研究旨在评估阻抗谱技术在诊断分娩过程中发生的括约肌损伤的适用性。这是一项前瞻性单中心研究。共纳入 22 名女性患者:10 名在产后早期出现括约肌功能障碍症状,12 名在自然分娩后 2 年以上的远期无影响。通过测量体格检查中的括约肌参数、肛门超声成像中的括约肌损伤程度和直肠测压中的括约肌功能参数,评估肛门括约肌的损伤情况。所有测量均作为参考,并与阻抗谱模型的结果进行比较。阻抗谱技术在与经肛门超声检查的比较中显示出最高的精度(平均准确率为 83.9%)。其结果有 74.1%与直肠体格检查的结果相对应,76.7%与直肠测压的结果相对应。该方法在评估括约肌的解剖和功能参数方面具有最高的准确性。新的阻抗谱技术有望用于检测产科肛门括约肌损伤。