For several purposes, skin parameters like thickness and elasticity can be measured. However, little is known about the accuracy of those measurements.
The aim of this study was to determine the intrarater and test-retest reliability of skin thickness and elasticity measurements performed with the DermaLab Combo®.
A total of 49 participants were included in this study. Skin thickness and elasticity were measured at six defined locations on the dominant arm. Measurements were repeated two times by the same observer to determine the test-retest reliability. To determine the inter-rater reliability, a second observer repeated the measurements once.
Inter-rater and test-retest reliability for elasticity measurements fluctuates per location and per parameter: Inter-rater intraclass correlation coefficient (ICC) ranged from 0.23 to 0.80, and test-retest ICC ranged from 0.25 to 0.84. Skin thickness was measured reliable by every observer on every location, with a test-retest ICC ranging from 0.71 to 0.83 and an inter-rater ICC ranging from 0.69 to 0.80.
The DermaLab Combo® showed a good inter-rater reliability when measuring skin thickness and elasticity. Not all locations are suitable for reliable inter-rater or test-retest measurements. The device is difficult to use by inexperienced users, as the echo probe is sensitive to small movements.
出于多种目的,可以测量皮肤厚度和弹性等参数。然而,人们对这些测量的准确性知之甚少。
本研究旨在确定使用 DermaLab Combo®进行皮肤厚度和弹性测量的组内和重测信度。
本研究共纳入 49 名参与者。在优势手臂的六个指定部位测量皮肤厚度和弹性。由同一位观察者重复测量两次以确定重测信度。为了确定观察者间信度,第二位观察者重复测量一次。
弹性测量的观察者间和重测信度因位置和参数而异:观察者间的组内相关系数(ICC)范围为 0.23 至 0.80,重测 ICC 范围为 0.25 至 0.84。每位观察者在每个部位都能可靠地测量皮肤厚度,重测 ICC 范围为 0.71 至 0.83,观察者间 ICC 范围为 0.69 至 0.80。
当测量皮肤厚度和弹性时,DermaLab Combo®显示出良好的观察者间信度。并非所有部位都适合进行可靠的观察者间或重测测量。对于没有经验的用户来说,该设备使用起来很困难,因为回声探头对小的移动很敏感。