J Neuromuscul Dis. 2019;6(2):233-240. doi: 10.3233/JND-190388.
Myotonic dystrophy type 1 (DM1) is the most prevalent degenerative neuromuscular disease in adults. Knee extensor (KE) maximal strength loss is a strong indicator of physical limitations in DM1. A reliable, precise and accessible maximal strength evaluation method needs to be validated for this slowly progressive disease.
This paper aims to assess the intra-rater reliability, the standard error of measurement (SEM), the minimal detectable change (MDC), and the concurrent validity of quantified muscle testing (QMT) using a handheld dynamometer with a gold standard: the Biodex isokinetic device.
Nineteen men with the adult form of DM1 participated in this study by attending 2 visits spaced by one week. The evaluation of KE muscle strength with QMT was completed on the first visit and the same QMT evaluation in addition to the maximal muscle strength evaluation using an isokinetic device were performed on the second visit.
The intra-rater reliability was excellent with an intraclass correlation coefficient (ICC) of 0.98 (0.96-0.99 : 95% confidence interval). SEM and MDC values were 1.05 Nm and 2.92 Nm, respectively. Concurrent validity of QMT of KE muscle group with the Biodex was also excellent with a Spearman's correlation of ρ= 0.98.
The excellent concurrent validity and intra-rater reliability, and the small SEM and MDC of the QMT make this test a method of choice, in either a clinical or research setting, to precisely evaluate muscle strength impairments of the KE in men with DM1.
1 型肌强直性营养不良(DM1)是成人中最常见的退行性神经肌肉疾病。伸膝肌(KE)最大力量损失是 DM1 患者身体受限的一个强有力指标。对于这种进展缓慢的疾病,需要验证一种可靠、精确且易于使用的最大力量评估方法。
本文旨在评估使用手持测力计进行定量肌肉测试(QMT)的组内可靠性、测量标准误差(SEM)、最小可检测变化(MDC)和同时效度,金标准是等速设备 Biodex。
19 名成年型 DM1 男性患者参加了这项研究,他们在间隔一周的 2 次就诊中进行了研究。第一次就诊时进行 KE 肌肉力量的 QMT 评估,第二次就诊时除了使用等速设备进行最大肌肉力量评估外,还进行了相同的 QMT 评估。
组内可靠性极好,组内相关系数(ICC)为 0.98(0.96-0.99:95%置信区间)。SEM 和 MDC 值分别为 1.05 Nm 和 2.92 Nm。KE 肌肉群 QMT 与 Biodex 的同时效度也极好,Spearman 相关系数 ρ=0.98。
QMT 的极好的同时效度和组内可靠性,以及较小的 SEM 和 MDC,使该测试成为一种选择,无论是在临床还是研究环境中,都可以精确评估 DM1 男性患者 KE 肌肉力量的损伤。