R2 relaxometry based MR imaging for estimation of liver iron content: A comparison between two methods.

机构信息

Division of Body Imaging, Department of Radiology, Children's Hospital of Philadelphia, 3401 Civic Center Blvd, Philadelphia, PA, 19104, USA.

Perelman School of Medicine, University of Pennsylvania, Philadelphia, PA, USA.

出版信息

Abdom Radiol (NY). 2019 Sep;44(9):3058-3068. doi: 10.1007/s00261-019-02074-4.

Abstract

PURPOSE

To compare the reproducibility and accuracy of R2-relaxometry MRI for estimation of liver iron concentration (LIC) between in-house analysis and FDA-approved commercially available third party results.

METHODS

All MR studies were performed on a 1.5T scanner. Multi-echo spin-echo scans with a fixed TR and increasing TE values of 6 ms, 9 ms, 12 ms, 15 ms, and 18 ms (spaced at 3 ms intervals) were used. Post-processing of the images to calculate mean relaxivity, R2, included drawing of regions of interest to include the whole liver on mid-slice. The relationship between liver R2 values and estimated LIC calculated with in-house analysis and values reported by an external company (FerriScan, Resonance Health, Australia) were assessed with correlation coefficients and Bland-Altman difference plots. Continuous variables are presented as mean ± standard deviation. Significance was set at p value < 0.05.

RESULTS

474 studies from 175 patients were included in the study (mean age 10.4 ± 4.2 years (range 1-18 years); 254 studies from girls, 220 studies from boys). LIC ranged from 0.6 to 43 mg/g dry tissue, covering a broad range from normal levels to extremely high iron levels. Linearity between proprietary and in-house methods was excellent across the observed range for R2 (31.5 to 334.8 s); showing a correlation coefficient of r = 0.87, p < 0.001. Bland-Altman R2 difference plot between the two methods shows a mean bias of + 21.5 s (range - 47.0 to + 90.0 s between two standard deviations). LIC reported by FerriScan compared with LIC estimated in-house with R2 as reported by FerriScan agreed strongly, (r = 1.0, p < 0.001).

CONCLUSION

R2 relaxometry MR imaging for liver iron concentration estimation is reproducible between proprietary FDA-approved commercial software and in-house analysis methods.

摘要

目的

比较内部分析和经 FDA 批准的商业第三方结果的 R2 弛豫率 MRI 对肝脏铁浓度(LIC)的估计的再现性和准确性。

方法

所有磁共振研究均在 1.5T 扫描仪上进行。使用多回波自旋回波扫描,固定 TR 和增加 TE 值分别为 6ms、9ms、12ms、15ms 和 18ms(间隔 3ms)。对图像进行后处理以计算平均弛豫率 R2,包括在中切片上绘制包括整个肝脏的感兴趣区。用相关系数和 Bland-Altman 差值图评估肝脏 R2 值与内部分析和外部公司(澳大利亚 Resonance Health 的 FerriScan)报告的估计 LIC 之间的关系。连续变量表示为平均值±标准差。显著性设定为 p 值<0.05。

结果

共纳入 175 例患者的 474 项研究(平均年龄 10.4±4.2 岁[范围 1-18 岁];254 项研究为女孩,220 项研究为男孩)。LIC 范围从 0.6 到 43mg/g 干组织,涵盖从正常水平到极高铁水平的广泛范围。在 R2 观察到的范围内,专用和内部方法之间的线性关系非常好(31.5 至 334.8s);显示相关系数 r=0.87,p<0.001。两种方法之间的 Bland-Altman R2 差值图显示平均偏差为+21.5s(两个标准差之间的范围为-47.0 至+90.0s)。FerriScan 报告的 LIC 与 FerriScan 报告的 R2 估计的内部 LIC 强烈一致(r=1.0,p<0.001)。

结论

R2 弛豫率 MRI 成像用于肝脏铁浓度估计在专有的经 FDA 批准的商业软件和内部分析方法之间具有可重复性。