Assessing the accuracy of CMRtools software for diagnosing liver iron overload in thalassemia patients: influencing factors and optimisation strategies.

作者信息

Luo Chaotian, Peng Fei, Xu Fengming, Tang Cheng, Zhang Yanyan, Huang Chaojie, Liang Linlin, Ning Xiaojing, Peng Peng

机构信息

Department of Radiology, The First Affiliated Hospital of Guangxi Medical University, Nanning, China.

NHC Key Laboratory of Thalassemia Medicine, Guangxi Medical University, Nanning, China.

出版信息

Front Med (Lausanne). 2024 Sep 20;11:1424294. doi: 10.3389/fmed.2024.1424294. eCollection 2024.

Abstract

BACKGROUND

CMRtools is a software package that can be used to measure T2* values to diagnose liver iron overload, however, its accuracy in terms is affected by multiple factors, including goodness-of-fit (R value), the number of echo time (TE) images, and the liver iron concentration (LIC). To investigate the effects of the R value, the number of TE images, and the LIC on the accuracy of CMRtools software for measuring T2* values to diagnose liver iron overload (LIO).

MATERIALS AND METHODS

CMRtools software was used to measure liver T2* values among 108 thalassemia patients via the truncation method, and the R values, the number of TE images, and T2* values were recorded. These values were subsequently converted into liver iron concentration (LIC) values. The LIC (derived from MRI-R2/FerriScan) was used as a reference, and the diagnostic accordance rate (DAR) was compared between R value subgroups, between TE image number subgroups, and between LIC subgroups.

RESULTS

The greater the R value was, the greater the standardized DAR (SDAR) was ( < 0.05). The SDAR are not identical between each TE image number subgroup ( > 0.05). However, the relationship between TE image number subgroups and SDAR was analysed using Spearman's correlation, and it was found to be positively correlated (  = 0.729,  = 0.017). The SDAR are not identical between each LIC subgroup ( > 0.05), furthermore, the relationship between LIC subgroup and SDAR was found irrelevant ( = 0.747).

CONCLUSION

The accuracy of CMRtools software for diagnosing LIO in patients with thalassemia can be improved by artificially controlling the number of TE images to be fitted and selecting higher R values.

摘要

背景

CMRtools是一个可用于测量T2值以诊断肝脏铁过载的软件包,然而,其在这方面的准确性受多种因素影响,包括拟合优度(R值)、回波时间(TE)图像数量和肝脏铁浓度(LIC)。为了研究R值、TE图像数量和LIC对CMRtools软件测量T2值诊断肝脏铁过载(LIO)准确性的影响。

材料与方法

使用CMRtools软件通过截断法测量108例地中海贫血患者的肝脏T2值,并记录R值、TE图像数量和T2值。随后将这些值转换为肝脏铁浓度(LIC)值。将来自MRI-R2/FerriScan的LIC用作参考,比较R值亚组、TE图像数量亚组和LIC亚组之间的诊断符合率(DAR)。

结果

R值越大,标准化诊断符合率(SDAR)越大(<0.05)。各TE图像数量亚组之间的SDAR不相同(>0.05)。然而,使用Spearman相关性分析TE图像数量亚组与SDAR之间的关系,发现呈正相关(=0.729,=0.017)。各LIC亚组之间的SDAR不相同(>0.05),此外,发现LIC亚组与SDAR之间的关系不相关(=0.747)。

结论

通过人工控制要拟合的TE图像数量并选择更高的R值,可以提高CMRtools软件诊断地中海贫血患者LIO的准确性。