J Rehabil Assist Technol Eng. 2016 May 5;3:2055668316642387. doi: 10.1177/2055668316642387. eCollection 2016 Jan-Dec.
Due to limited accuracy of self-reported activities of lower limb prosthetic users, there is increasing interest in providing accurate walking time for those who are mobilising using their prosthesis in the rehabilitation ward. The aim of this study was to test the accuracy of a tri axial accelerometer (ActivPAL) in measuring walking activity of amputee patients while using their prosthetic limb.
For the study, 21 subjects wore accelerometer devices taped to the thigh on both the amputated and non-amputated side. Each subject was asked to perform a set of activity routines: walking with prosthesis for 5 minutes, self-propelling in a wheel chair for 3 minutes and being pushed in a wheel chair for 3 minutes. Each activity was observed and timed by a physiotherapist and the observed times were compared with the output from accelerometer monitoring.
Using the Bland-Altman method, the mean difference between observed and ActivPal monitor for total time spent walking for the non-amputated side was 0.004 seconds (limits of agreement -0.09 to 0.10 seconds) and for amputated side was 0.11 seconds (limits of agreement -0.43 to 0.66 seconds). An analysis between monitored and observed time found the sensitivity for the non-amputated side was 90.5% and 86% for the amputated side.
The use of ActivPAL accelerometers in measuring walking time for an individual using a prosthesis is within acceptable levels of accuracy for continuous ambulation monitoring. It has potentially important clinical application for prescribing prostheses.
由于下肢假肢使用者自我报告活动的准确性有限,人们越来越关注为在康复病房使用假肢活动的患者提供准确的行走时间。本研究的目的是测试三轴加速度计(ActivPAL)在测量截肢患者使用假肢时行走活动的准确性。
在本研究中,21名受试者在截肢侧和非截肢侧的大腿上都佩戴了粘贴式加速度计设备。要求每名受试者执行一组活动程序:使用假肢行走5分钟、在轮椅上自行推动3分钟以及被推着坐在轮椅上3分钟。每项活动由一名物理治疗师进行观察并计时,并将观察到的时间与加速度计监测的输出结果进行比较。
使用Bland-Altman方法,非截肢侧行走总时间的观察值与ActivPal监测值之间的平均差异为0.004秒(一致性界限为-0.09至0.10秒),截肢侧为0.11秒(一致性界限为-0.43至0.66秒)。对监测时间和观察时间的分析发现,非截肢侧的敏感度为90.5%,截肢侧为86%。
在测量使用假肢的个体的行走时间时,使用ActivPAL加速度计的准确性在连续步行监测的可接受水平内。它在假肢处方方面具有潜在的重要临床应用价值。