评估一种便携式离子选择性电极仪测量牛全血、血浆、牛奶、皱胃液和尿液中钠离子浓度的分析性能。

Evaluation of the analytical performance of a portable ion-selective electrode meter for measuring whole-blood, plasma, milk, abomasal-fluid, and urine sodium concentrations in cattle.

机构信息

Department of Veterinary Clinical Medicine, College of Veterinary Medicine, University of Illinois at Urbana-Champaign 61802; Department of Animal Medicine (Internal Medicine), Faculty of Veterinary Medicine, Benha University Moshtohor-Toukh, Kalyobiya, Egypt 13736.

Department of Veterinary Clinical Studies, Faculty of Veterinary Medicine, Universiti Putra, Selangor, Malaysia 43400.

出版信息

J Dairy Sci. 2019 Aug;102(8):7435-7444. doi: 10.3168/jds.2018-16198. Epub 2019 Jun 13.

Abstract

A portable ion-selective electrode (ISE) meter (LAQUAtwin B-722; Horiba Instruments Inc., Irvine, CA) is available for measuring the sodium ion concentration ([Na]) in biological fluids. The objective of this study was to characterize the analytical performance of the ISE meter in measuring [Na] in whole-blood, plasma, milk, abomasal fluid, and urine samples from cattle. Method comparison studies were performed using whole-blood and plasma samples from 106 sick calves and 11 sick cows admitted to a veterinary teaching hospital, 80 milk and 206 urine samples from 16 lactating Holstein-Friesian cows with experimentally induced free water, electrolyte, and acid-base imbalances, and 67 abomasal fluid samples from 7 healthy male Holstein-Friesian calves fed fresh milk with or without an oral electrolyte solution. Deming regression and Bland-Altman plots were used to determine the accuracy of the meter against reference methods. The meter used in direct mode on undiluted samples measured whole-blood [Na] 9.7 mmol/L (7.3%) lower than a direct ISE reference method and plasma [Na] 16.7 mmol/L (12.7%) lower than an indirect ISE reference method. The meter run in direct mode measured milk [Na] 3.1 mmol/L lower and abomasal fluid [Na] 9.0% lower than indirect ISE reference methods. The meter run in indirect mode on diluted samples accurately measured urine [Na] compared with an indirect ISE reference method. We conclude that, after adjustment for the bias determined from Bland-Altman plots, the LAQUAtwin ISE meter provides a clinically useful and low-cost cow-side instrument for measuring [Na] in whole blood, plasma, milk, and abomasal fluid.

摘要

一种便携式离子选择性电极(ISE)计(LAQUAtwin B-722;Horiba 仪器公司,欧文,加利福尼亚州)可用于测量生物体液中的钠离子浓度([Na])。本研究的目的是表征 ISE 计在测量牛的全血、血浆、牛奶、皱胃液和尿液样本中的[Na]的分析性能。使用来自兽医教学医院的 106 头患病小牛和 11 头患病奶牛的全血和血浆样本、80 份牛奶和 206 份尿液样本以及 7 头健康的荷斯坦-弗里生雄性小牛的 67 份皱胃液样本进行了方法比较研究,这些小牛喂食了新鲜牛奶和/或口服电解质溶液,导致自由水、电解质和酸碱平衡紊乱。使用 Deming 回归和 Bland-Altman 图确定了该计与参考方法的准确性。直接模式下未经稀释的样本测量全血[Na]低 9.7mmol/L(7.3%),比直接 ISE 参考方法;血浆[Na]低 16.7mmol/L(12.7%),比间接 ISE 参考方法。直接模式下,该计测量牛奶[Na]低 3.1mmol/L,皱胃液[Na]低 9.0%,比间接 ISE 参考方法。间接模式下对稀释样本的测量结果与间接 ISE 参考方法相比准确测量了尿液[Na]。我们得出结论,经 Bland-Altman 图确定的偏差调整后,LAQUAtwin ISE 计为临床提供了一种有用且低成本的牛场仪器,可用于测量全血、血浆、牛奶和皱胃液中的[Na]。