因空化作用导致机械主动脉瓣假体上血栓的超声心动图模拟物:压力恢复的矛盾现象。

Echocardiographic mimicker of thrombus on a mechanical aortic valve prosthesis due to cavitation: A paradoxical phenomenon of pressure recovery.

作者信息

Wang Zhenzhen, Ayoub Chadi, Thaden Jeremy J, Alsidawi Said, Miller Fletcher A, Sinak Lawrence J, Rihal Charanjit S, Melduni Rowlens M

机构信息

Department of Cardiovascular Medicine, Mayo Clinic, Rochester, Minnesota.

出版信息

Echocardiography. 2019 Jul;36(7):1397-1400. doi: 10.1111/echo.14403. Epub 2019 Jun 17.

Abstract

We describe a case of a mass-like echocardiographic density on a mechanical prosthetic aortic valve. We initially suspected a thrombus vs vegetation on transthoracic echocardiography, but after transesophageal echocardiography, the density was subsequently determined to be cavitation by reviewing the initial images in slow motion.

摘要

我们描述了一例机械人工主动脉瓣上出现类似肿块的超声心动图致密影的病例。经胸超声心动图检查时,我们最初怀疑是血栓与赘生物,但经食管超声心动图检查后,通过慢动作回顾初始图像,随后确定该致密影为空洞形成。