Department of Ophthalmology and Graduate School of Medicine and Pharmaceutical Sciences, University of Toyama, Toyama, Japan.
J Glaucoma. 2019 Sep;28(9):808-810. doi: 10.1097/IJG.0000000000001318.
We measured changes in the corneal morphology and intraocular pressure (IOP) before and after the use of contact lens sensor (CLS). Then, we investigated whether the ocular dimensional change affected the measured IOP value with CLS.
In this prospective study, we examined 112 eyes of glaucoma patients. We measured the IOP fluctuation using a Sensimed Triggerfish CLS. We measured the corneal morphology before and after making measurements using a CLS in anterior segment optic coherent optical coherence tomography and measured the IOP with a Goldmann applanation tonometer. We investigated the correlation between these ocular dimensional changes and the last IOP value with CLS.
Measurement with the CLS significantly increased the IOP (P=0.0305), made the corneal thickness (CT) significantly thicker (P<0.001), and increased the steeper meridian (P=0.0026). The last value of IOP-patterns recorded with CLS was significantly correlated with the IOP immediately after the measurement with Goldmann applanation tonometer (P=0.0134), was not significantly correlated with central and peripheral CT. The difference of steeper and flatter curvature between before and after measurement significantly correlated with the last value of IOP-patterns recorded with CLS.
Measurement with a CLS affects the IOP, CT, and corneal steeper meridian. From the last value of IOP-related patterns recorded with CLS, CLS could perceive accurate IOP. The change of corneal curvature might affect CLS output values.
我们测量了使用接触镜传感器 (CLS) 前后角膜形态和眼内压 (IOP) 的变化,然后研究了眼球尺寸变化是否会影响使用 CLS 测量的 IOP 值。
在这项前瞻性研究中,我们检查了 112 只青光眼患者的眼睛。我们使用 Sensimed Triggerfish CLS 测量 IOP 波动。我们使用眼前节光学相干光断层扫描在使用 CLS 进行测量之前和之后测量角膜形态,并使用 Goldmann 压平眼压计测量 IOP。我们研究了这些眼球尺寸变化与最后一次使用 CLS 测量的 IOP 值之间的相关性。
使用 CLS 测量会显著增加 IOP(P=0.0305),使角膜厚度(CT)显著增厚(P<0.001),并增加陡峭子午线(P=0.0026)。CLS 记录的最后一次 IOP 模式值与 Goldmann 压平眼压计测量后的即刻 IOP 值显著相关(P=0.0134),与中央和周边 CT 无显著相关性。测量前后陡峭曲率和扁平曲率的差异与 CLS 记录的最后一次 IOP 模式值显著相关。
CLS 测量会影响 IOP、CT 和角膜陡峭子午线。从 CLS 记录的最后一次与 IOP 相关的模式值可以感知到准确的 IOP。角膜曲率的变化可能会影响 CLS 的输出值。