Acta Ophthalmol. 2012 Nov;90(7):e534-9. doi: 10.1111/j.1755-3768.2012.02495.x. Epub 2012 Sep 13.
To assess the effect of overnight wear of a contact lens-based sensor (CLS) for monitoring of 24-hr intraocular pressure (IOP) fluctuations on central corneal thickness (CCT).
Changes in the CCT, mid-peripheral corneal thickness and central corneal radius (CCR) during overnight CLS wear in 20 eligible patients with ocular hypertension or established glaucoma were prospectively studied using ultrasound pachymetry and topography. Corneal thickness and CCR changes were evaluated from pre-to-postsleep, with the fellow eye as control. Paired t-test or Wilcoxon signed-rank test was used as appropriate and with α = 0.05. Relationship between the IOP profile recorded by the CLS and the pre-to-postsleep corneal thickness differences was assessed using the Spearman correlation coefficient.
After CLS wear, mean CCT had changed from 523 to 537 μm (p = 0.015) in the study eye and from 518 to 522 (p = 0.206) in the fellow eye (n = 15). There was no difference in CCT change between eyes (p = 0.075). There were no statistically significant changes in horizontal or vertical CCR in either eye (p > 0.05 for all). No correlation was found between the pre-to-postsleep differences in the CLS signal and the pre-to-postsleep differences in ultrasound CCT measurements (p = 0.974).
The continuous IOP monitoring does not appear to be affected by differences in corneal thickness that occur during overnight CLS wear, although the CLS did induce some corneal swelling. This effect was not statistically significantly different from the control eye and does not seem to influence the CLS IOP profile.
评估用于监测 24 小时眼压(IOP)波动的接触镜式传感器(CLS)的过夜佩戴对中央角膜厚度(CCT)的影响。
使用超声角膜测厚仪和角膜地形图对 20 名患有高眼压或已确诊青光眼的患者的眼角膜厚度(CCT)、中周边角膜厚度和中央角膜半径(CCR)在夜间 CLS 佩戴期间的变化进行前瞻性研究。使用配对 t 检验或 Wilcoxon 符号秩检验评估睡眠前后的角膜厚度和 CCR 变化,以对侧眼作为对照。使用 Spearman 相关系数评估 CLS 记录的 IOP 曲线与睡眠前后角膜厚度差异之间的关系。
在 CLS 佩戴后,研究眼的 CCT 从 523μm 增加到 537μm(p=0.015),对侧眼从 518μm 增加到 522μm(p=0.206)(n=15)。双眼之间的 CCT 变化无差异(p=0.075)。在两眼的水平或垂直 CCR 均无统计学显著变化(p>0.05)。CLS 信号的睡眠前后差异与超声 CCT 测量的睡眠前后差异之间未发现相关性(p=0.974)。
尽管 CLS 确实会引起一些角膜肿胀,但连续的 IOP 监测似乎不受夜间 CLS 佩戴期间发生的角膜厚度差异的影响。这种影响与对照眼无统计学显著差异,似乎不会影响 CLS 的 IOP 曲线。