Pietersz R N, Reesink H W, Dekker W J
Red Cross Blood Bank Amsterdam, The Netherlands.
Vox Sang. 1987;53(4):208-13. doi: 10.1111/j.1423-0410.1987.tb05068.x.
Six citrate phosphate dextrose (CPD)-saline adenine glucose mannitol (SAG M) quadruple systems were evaluated for the preparation and storage of leukocyte-poor platelet concentrates (PC) from buffy coats. The platelet storage bags examined were manufactured from normal polyvinylchloride (PVC) or special-type plastics. Biotest supplied PVC 76 (n = 14) and PVC 763 (n = 16) NPBI supplied PSV 3277 (n = 15) and DPL-110 (n = 14) and Terumo supplied Teruflex (n = 18) and molded Teruflex (n = 14). The PC were stored for 7 days at 22 degrees C on a horizontally shaking platform. Cell counts, pH, PO2, PCO2, morphology score and swirling patterns were monitored at 5, 72, 120 and 168 h. The plasma volumes averaged 63 ml and ranged from 39 to 81 ml, the overall mean +/- SD platelet concentration was 0.89 +/- 0.33 X 10(9)/ml. None of the PC had a leukoyte count higher than 10 X 10(6) per unit. After storage for 168 h, the pH ranged from 6.56 to 7.40 for all brands. The PO2 remained stable and even rose significantly (p less than 0.05) during storage in the NPBI PSV 3277 and Terumo molded Teruflex bags. The PCO2 decreased equally in all bags. Morphology scores were well maintained in 98% of all PC for up to 120 h, and in 83% at 168 h. A swirling pattern score of 2.5 or greater predicted with a specificity of 100% a good morphology score in the PC.(ABSTRACT TRUNCATED AT 250 WORDS)
对六个枸橼酸盐-磷酸盐-葡萄糖(CPD)-生理盐水-腺嘌呤-葡萄糖-甘露醇(SAG-M)四联体系进行了评估,以用于从 Buffy 层制备和储存少白细胞血小板浓缩物(PC)。所检查的血小板储存袋由普通聚氯乙烯(PVC)或特殊类型塑料制成。Biotest 提供了 PVC 76(n = 14)和 PVC 763(n = 16);NPBI 提供了 PSV 3277(n = 15)和 DPL - 110(n = 14);Terumo 提供了 Teruflex(n = 18)和模制 Teruflex(n = 14)。将 PC 在 22℃的水平振荡平台上储存 7 天。在 5、72、120 和 168 小时监测细胞计数、pH、PO₂、PCO₂、形态学评分和漩涡模式。血浆体积平均为 63ml,范围为 39 至 81ml,总体平均±标准差血小板浓度为 0.89±0.33×10⁹/ml。所有 PC 的白细胞计数均不高于每单位 10×10⁶。储存 168 小时后,所有品牌的 pH 范围为 6.56 至 7.40。在 NPBI PSV 3277 和 Terumo 模制 Teruflex 袋中储存期间,PO₂保持稳定,甚至显著升高(p<0.05)。所有袋子中的 PCO₂均同等程度下降。在所有 PC 中,98%在长达 120 小时内形态学评分保持良好,168 小时时为 83%。漩涡模式评分为 2.5 或更高时,预测 PC 中良好形态学评分的特异性为 100%。(摘要截短于 250 字)