超声测量产后女性膀胱容量的有效性。

Validity of bladder volume measurement by ultrasound in women postpartum.

机构信息

Department of Obstetrics and Gynecology, Herlev Hospital, Herlev and Gentofte University Hospital, Copenhagen, Denmark.

出版信息

Int Urogynecol J. 2020 Mar;31(3):643-649. doi: 10.1007/s00192-019-04037-1. Epub 2019 Jul 5.

Abstract

INTRODUCTION

Vaginal birth increases the risk of urinary retention and overdistention of the bladder. To avoid unnecessary discomfort by catheterization, it is preferable to use ultrasound for diagnosis of these conditions. The aim of this study was to determine the validity of transabdominal ultrasound and a portable ultrasound system, Biocon-700, to measure bladder volume in women postpartum.

METHODS

Fifty women were included in this method comparison study. Within 48 h of giving birth, their bladder volume was measured in randomized order with both transabdominal ultrasound and Biocon-700. After urination, participants were scanned with Biocon-700 to measure residual bladder volume, and finally the bladder was emptied by catheterization. The total bladder volume was calculated as the voided volume plus the catheterized volume.

RESULTS

Biocon-700 measured 43.4 ml ± 117.4 ml (mean ± SD) lower than the total bladder volume, while volumes measured by transabdominal ultrasound were 117.8 ml ± 110.0 ml (mean ± SD) lower. Linear regression showed significant proportional bias in both methods. The Biocon-700 could detect a residual bladder volume > 150 ml with a positive predictive value of 66.7% and a negative predictive value of 100%.

CONCLUSIONS

Neither transabdominal ultrasound nor the portable ultrasound system, Biocon-700, can be used to measure bladder volume precisely after vaginal delivery. However, both ultrasound methods can be used as screening tools to prevent overdistention of the bladder, and Biocon-700 can furthermore be used to screen women for a residual bladder volume > 150 ml.

摘要

介绍

阴道分娩会增加尿潴留和膀胱过度膨胀的风险。为了避免因导尿引起的不必要的不适,最好使用超声来诊断这些情况。本研究旨在确定经腹超声和便携式超声系统 Biocon-700 测量产后妇女膀胱容量的有效性。

方法

本方法比较研究纳入了 50 名女性。在分娩后 48 小时内,以随机顺序使用经腹超声和 Biocon-700 测量她们的膀胱容量。排尿后,使用 Biocon-700 扫描以测量残余膀胱容量,最后通过导尿排空膀胱。膀胱总容量为排空量加导尿量。

结果

Biocon-700 测量的膀胱总容量比实际值低 43.4 毫升±117.4 毫升(均值±标准差),而经腹超声测量的容量则低 117.8 毫升±110.0 毫升。线性回归显示两种方法均存在显著的比例偏差。Biocon-700 可检测残余膀胱容量>150 毫升,阳性预测值为 66.7%,阴性预测值为 100%。

结论

经腹超声和便携式超声系统 Biocon-700 均不能准确测量阴道分娩后的膀胱容量。然而,这两种超声方法都可以用作筛查工具,以防止膀胱过度膨胀,而 Biocon-700 还可以用于筛查残余膀胱容量>150 毫升的女性。