Department of Radiology, Women's Imaging Unit, National Cancer Institute, Cairo University, Cairo, Egypt.
Department of Radiology, Women's Imaging Unit, Kasr ElAiny Hospital, Cairo University, Cairo, Egypt.
Clin Radiol. 2019 Oct;74(10):771-781. doi: 10.1016/j.crad.2019.06.002. Epub 2019 Jul 6.
To assess the utility of contrast-enhanced spectral mammography (CESM) to detect malignancy in the postoperative breast of cancer patients.
The study was a prospective analysis of 76 women following operations for breast cancer. CESM was performed with low (22-33 kVp) and high (44-49 kVp) energy exposures after intravenous injection of contrast agent. Breast abnormalities were evaluated regarding the presence, morphology, and pattern of contrast medium uptake. The reference standard was histopathology of the Tru-Cut biopsy or surgical biopsy in cases where malignancy was suspected and absence of abnormality at follow-up sessions in benign cases.
Seventy cases were eligible for data analysis, and malignancy was detected in 48.6% (n=34/70). Enhancement at the operative bed was observed in 40 (57.14%) lesions. At traditional mammography, the diagnosis was false negative in 17/70 and false positive in 28/70 lesions, and presented a low sensitivity of 50%, a very low specificity of 22%, a positive predictive value of 37.7%, a negative predicative value of 32%, and an accuracy of 35.7%. CESM enhanced the performance of mammography and produced a sensitivity of 91.17%, a specificity of 75%, a positive predictive value of 77.5%, a negative predictive value of 90%, and an accuracy of 82.85%.
CESM is a credible technique that could be used in conjunction with the traditional mammogram to screen for cancer in the postoperative breast.
评估对比增强光谱乳腺摄影术(CESM)在检测乳腺癌术后乳房恶性肿瘤中的应用价值。
这是一项对 76 例乳腺癌术后患者的前瞻性分析研究。在静脉注射造影剂后,采用低能(22-33 kVp)和高能(44-49 kVp)进行 CESM。根据造影剂摄取的存在、形态和模式评估乳腺异常。参考标准是恶性病变怀疑时的 tru-cut 活检或手术活检的组织病理学,良性病例的随访无异常。
70 例病例符合数据分析条件,48.6%(n=34/70)检测到恶性肿瘤。40 个病变(57.14%)在手术床处有增强。在传统乳腺摄影中,70 个病变中有 17 个(17/70)为假阴性,28 个(28/70)为假阳性,其灵敏度为 50%,特异性为 22%,阳性预测值为 37.7%,阴性预测值为 32%,准确率为 35.7%。CESM 增强了乳腺摄影的性能,其灵敏度为 91.17%,特异性为 75%,阳性预测值为 77.5%,阴性预测值为 90%,准确率为 82.85%。
CESM 是一种可靠的技术,可与传统乳腺 X 线摄影联合用于筛查乳腺癌术后乳房的癌症。