Author to whom any correspondence should be addressed.
Phys Med Biol. 2019 Aug 28;64(17):175003. doi: 10.1088/1361-6560/ab3317.
The step-and-shoot method of pencil beam scanning delivers the dose on a 3D grid in the target volume, with one dimension defined by the proton energy. While the dose per pencil beam may vary substantially within an iso-energy layer, the beam current typically remains constant. In this static operation mode, the inherent latency of the beam switch-off mechanism results in a lower limit for the deliverable spot dose, which may prevent the application of some of the low-weighted spots prescribed by the treatment planning system. To overcome this limitation, we introduced dynamic beam current control at the PSI Gantry 2, an innovative new approach successfully commissioned and in clinical operation since fall 2017. The control system was enhanced with a direct link to the vertical deflector located at the centre of the cyclotron. This connection allows much faster beam current changes (~0.1 ms) and hence opens up the possibility of dynamically reducing the current for individual low-dose spots. We demonstrate that with this new dynamic operation mode, all spots are delivered as planned without compromising treatment time. We show by two independent and complementary methods that the delivered dose distribution is improved.
笔形束扫描的步进-射击方法在靶区的 3D 网格上递送剂量,其中一个维度由质子能量定义。虽然每个笔形束的剂量在等能量层内可能会有很大差异,但束流通常保持不变。在这种静态操作模式下,束关闭机制的固有延迟导致可递送电喷雾剂量的下限,这可能会阻止治疗计划系统规定的一些低权重点的应用。为了克服这一限制,我们在 PSI 龙门架 2 上引入了动态束流控制,这是一种创新的新方法,自 2017 年秋季以来成功委托并投入临床使用。控制系统通过与位于回旋加速器中心的垂直偏转器的直接连接进行了增强。这种连接允许更快的束流变化(~0.1 ms),从而为单个低剂量点动态降低电流成为可能。我们证明,通过这种新的动态操作模式,可以按计划递送电喷雾,而不会影响治疗时间。我们通过两种独立且互补的方法表明,所递送电喷雾剂量分布得到了改善。