便携式呼吸分析仪(AIRE)评估乳糖吸收不良的有效性。

Validity of a Portable Breath Analyser (AIRE) for the Assessment of Lactose Malabsorption.

机构信息

The Liggins Institute, The University of Auckland, 85 Park Road, Grafton, Auckland 1023, New Zealand.

The Riddet Institute, Palmerston North 4442, New Zealand.

出版信息

Nutrients. 2019 Jul 17;11(7):1636. doi: 10.3390/nu11071636.

Abstract

Hydrogen (H) measurement in exhaled breath is a reliable and non-invasive method to diagnose carbohydrate malabsorption. Currently, breath H measurement is typically limited to clinic-based equipment. A portable breath analyser (AIRE, FoodMarble Digestive Health Limited, Dublin, Ireland) is a personalised device marketed for the detection and self-management of food intolerances, including lactose malabsorption (LM). Currently, the validity of this device for breath H analysis is unknown. Individuals self-reporting dairy intolerance (six males and six females) undertook a lactose challenge and a further seven individuals (all females) underwent a milk challenge. Breath samples were collected prior to and at frequent intervals post-challenge for up to 5 h with analysis using both the AIRE and a calibrated breath hydrogen analyser (BreathTracker, QuinTron Instrument Company Inc., Milwaukee, WI, USA). A significant positive correlation ( < 0.001, r > 0.8) was demonstrated between AIRE and BreathTracker H values, after both lactose and milk challenges, although 26% of the AIRE readings demonstrated the maximum score of 10.0 AU. Based on our data, the cut-off value for LM diagnosis (25 ppm H) using AIRE is 3.0 AU and it is effective for the identification of a response to lactose-containing foods in individuals experiencing LM, although its upper limit is only 81 ppm.

摘要

呼气中氢气(H)的测量是诊断碳水化合物吸收不良的一种可靠且非侵入性的方法。目前,呼气 H 测量通常限于临床设备。一种便携式呼吸分析仪(AIRE,FoodMarble Digestive Health Limited,都柏林,爱尔兰)是一种针对食物不耐受(包括乳糖吸收不良)的检测和自我管理而销售的个性化设备。目前,该设备用于呼气 H 分析的有效性尚不清楚。自我报告有乳制品不耐受的个体(六名男性和六名女性)接受了乳糖挑战,另有七名个体(均为女性)接受了牛奶挑战。在挑战前和挑战后频繁的时间点采集呼气样本,最多持续 5 小时,使用 AIRE 和校准的呼吸氢气分析仪(BreathTracker,QuinTron Instrument Company Inc.,密尔沃基,WI,美国)进行分析。在乳糖和牛奶挑战后,AIRE 和 BreathTracker 的 H 值之间显示出显著的正相关(<0.001,r>0.8),尽管 AIRE 的读数中有 26%显示出最大分数 10.0 AU。根据我们的数据,使用 AIRE 诊断乳糖吸收不良的截断值(25 ppm H)为 3.0 AU,它可有效识别乳糖不耐受个体对含乳糖食物的反应,尽管其上限仅为 81 ppm。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/af2a/6683064/31d6b2a30462/nutrients-11-01636-g001.jpg

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验