Ogawa Ken, Koyama Shouhei, Haseda Yuuki, Fujita Keiichi, Ishizawa Hiroaki, Fujimoto Keisaku
Graduate School of Science and Technology, Shinshu University, 3-15-1 Tokida, Ueda, Nagano 386-8567, Japan.
Research & Development Sec. Development Center, Naganokeiki Co., Ltd., 2150 Ikuta, Ueda, Nagano 386-0411, Japan.
Sensors (Basel). 2019 Jul 22;19(14):3222. doi: 10.3390/s19143222.
A small-size, high-precision fiber Bragg grating interrogator was developed for continuous plethysmograph monitoring. The interrogator employs optical edge filters, which were integrated with a broad-band light source and photodetector to demodulate the Bragg wavelength shift. An amplifier circuit was designed to effectively amplify the plethysmograph signal, obtained as a small vibration of optical power on the large offset. The standard deviation of the measured Bragg wavelength was about 0.1 pm. The developed edge filter module and amplifier circuit were encased with a single-board computer and communicated with a laptop computer via Wi-Fi. As a result, the plethysmograph was clearly obtained remotely, indicating the possibility of continuous vital sign measurement.