Suppr超能文献

[用于测定总补体活性及C3和C4组分的Optilite分析仪的评估]

[Evaluation of the Optilite analyser for determination of total complement activity and C3 and C4 fractions].

作者信息

Nespola Benoit, Comitogianni Hélène, Jahn Isabelle, Goetz Joëlle

机构信息

Laboratoire d'immunologie, Hôpitaux Universitaires de Strasbourg, France.

出版信息

Ann Biol Clin (Paris). 2019 Aug 1;77(4):447-452. doi: 10.1684/abc.2019.1453.

Abstract

The complement system is composed of a set of plasma or membrane proteins. Complement protein deficiencies can be inherited or acquired, through the presence of autoantibodies or through consumption. We evaluated the analytical performance of the Optilite analyser for the determination of the C3 and C4 levels and for the evaluation of the total complement activity. The intra- and inter-series CVs were evaluated and have showed satisfactory results, the concordances with analysers currently used in the laboratory (BNII and BCT, Siemens) are very good, as is the agreement between the serum and plasma samples. We also determined the reference values for the different parameters tested in view of a routine use of Optilite analyser in the laboratory.

摘要

补体系统由一组血浆或膜蛋白组成。补体蛋白缺乏症可以是遗传性的,也可以是后天获得的,其原因包括自身抗体的存在或补体的消耗。我们评估了Optilite分析仪在测定C3和C4水平以及评估总补体活性方面的分析性能。评估了批内和批间变异系数,结果令人满意,与实验室目前使用的分析仪(西门子公司的BNII和BCT)的一致性非常好,血清和血浆样本之间的一致性也很好。鉴于Optilite分析仪将在实验室常规使用,我们还确定了所测试的不同参数的参考值。