Guangdong Province Key Laboratory of Display Material and Technology, State Key Laboratory of Optoelectronic Materials and Technologies, School of Electronics and Information Technology, Sun Yat-sen University, No. 132 East Waihuan Road, Guangzhou 510006, China.
Sun Yat-sen University Shunde Research Institute, No. 9 Eastern Nanguo Road, Shunde District, Foshan 523800, China.
Sensors (Basel). 2019 Jul 23;19(14):3241. doi: 10.3390/s19143241.
In this work, we report an active respiration monitoring sensor based on a piezoelectric-transducer-gated thin-film transistor (PTGTFT) aiming to measure respiration-induced dynamic force in real time with high sensitivity and robustness. It differs from passive piezoelectric sensors in that the piezoelectric transducer signal is rectified and amplified by the PTGTFT. Thus, a detailed and easy-to-analyze respiration rhythm waveform can be collected with a sufficient time resolution. The respiration rate, three phases of respiration cycle, as well as phase patterns can be further extracted for prognosis and caution of potential apnea and other respiratory abnormalities, making the PTGTFT a great promise for application in long-term real-time respiration monitoring.
在这项工作中,我们报告了一种基于压电换能器门控薄膜晶体管(PTGTFT)的主动呼吸监测传感器,旨在实时测量呼吸引起的动态力,具有高灵敏度和鲁棒性。它与无源压电传感器的不同之处在于,压电换能器信号通过 PTGTFT 进行整流和放大。因此,可以收集到详细且易于分析的呼吸节律波形,并且具有足够的时间分辨率。进一步可以提取呼吸率、呼吸周期的三个阶段以及相位模式,以预测和警惕潜在的呼吸暂停和其他呼吸异常,这使得 PTGTFT 在长期实时呼吸监测中有很好的应用前景。