Schwind J, Gross M
Institut für Phasiologie der Friedrich-Schiller-Universität Jena.
EEG EMG Z Elektroenzephalogr Elektromyogr Verwandte Geb. 1988 Jun;19(2):96-100.
The analysis of electroencephalograms and event-related potentials may be difficult by the presence of eyelid and eye movement artifacts. In this paper a method is described, which allows a relatively simple on-line correction of vertical ocular activity, measurable by electro-oculogram, in several EEG channels. Eyelid and eye movement artifacts are being manifest in a different extension at a different EEG recording position. Using the electrooculogram, with the aid of a simple electronic device it is possible to differ between eyelid and eye movement artifacts, which will be corrected selectively. If only midline derivations are evaluated (Fz, Cz, Pz) the vertical EOG is sufficient and horizontal EOG may be left out from the correcting procedure. The efficiency of the procedure is proved by EEG and ERP records.
由于存在眼睑和眼球运动伪迹,脑电图和事件相关电位的分析可能会很困难。本文描述了一种方法,该方法可以对几个脑电图通道中通过眼电图测量的垂直眼球活动进行相对简单的在线校正。眼睑和眼球运动伪迹在不同的脑电图记录位置有不同程度的表现。借助一个简单的电子设备,利用眼电图可以区分眼睑和眼球运动伪迹,并对其进行选择性校正。如果只评估中线导联(Fz、Cz、Pz),垂直眼电图就足够了,水平眼电图可以在校正过程中省略。脑电图和事件相关电位记录证明了该方法的有效性。