Moudi Ehsan, Haghanifar Sina, Johari Maryam, Gholinia Hemmat, Ghanbarabadi Mohammad Kazemi
Oral Health Research Center, Institute of Health, Babol, Iran.
Dental Materials Research Center, Health Research Institute, Babol, Iran.
J Indian Soc Periodontol. 2019 Jul-Aug;23(4):334-338. doi: 10.4103/jisp.jisp_675_18.
Due to the extensive use of cone-beam computed tomography (CBCT) in dentistry, especially in measuring thicknesses of hard and soft tissues, and the fact that CBCT has low contrast resolution, the aim of this study was to evaluate the accuracy of CBCT in measuring soft tissue thickness.
In this cross-sectional study, pink baseplate wax as a soft tissue simulation was overlaid in different areas of the jaws on the dry human skull, and the probe was used to determine the thicknesses of 1, 2, 3, and 5 mm. These thicknesses were then measured accurately with a digital caliper by one person four times at interval of a week. The skull was scanned using two CBCT scanners, and the different thicknesses of wax were measured by two observers four times with a week interval. The CBCT measurements were compared with digital caliper measurements as a reference.
Statistical analysis showed no significant difference between CBCT and digital caliper measurements for thickness < 2 mm ( > 0.05).
Statistically, the difference between digital caliper and CBCT measurements was < 0.1 mm, but this difference is not clinically important. The accuracy of CBCT in measuring soft tissue thickness was 0.1 mm.
由于锥形束计算机断层扫描(CBCT)在牙科中的广泛应用,尤其是在测量硬组织和软组织厚度方面,且CBCT的对比分辨率较低,本研究旨在评估CBCT测量软组织厚度的准确性。
在这项横断面研究中,将粉色基板蜡作为软组织模拟物覆盖在干燥人头骨的颌骨不同区域,并用探针确定1、2、3和5毫米的厚度。然后由一人用数字卡尺以一周为间隔对这些厚度进行四次精确测量。使用两台CBCT扫描仪对颅骨进行扫描,由两名观察者以一周为间隔对蜡的不同厚度进行四次测量。将CBCT测量结果与作为参考的数字卡尺测量结果进行比较。
统计分析表明,对于厚度<2毫米的情况,CBCT测量结果与数字卡尺测量结果之间无显著差异(>0.05)。
从统计学角度来看,数字卡尺与CBCT测量结果之间的差异<0.1毫米,但这种差异在临床上并不重要。CBCT测量软组织厚度的准确性为0.1毫米。