Suppr超能文献

使用干燥滤纸样本及气相色谱-串联质谱法(GC-MS/MS)评估尿雌激素和孕酮代谢物。

Evaluating urinary estrogen and progesterone metabolites using dried filter paper samples and gas chromatography with tandem mass spectrometry (GC-MS/MS).

作者信息

Newman Mark, Pratt Suzanne M, Curran Desmond A, Stanczyk Frank Z

机构信息

Precision Analytical Inc., 3138 NE Rivergate Street #301C, McMinnville, OR 97128 USA.

Suzanne Pratt Works LLC, 7114 SE 18th Avenue, Portland, OR 97202 USA.

出版信息

BMC Chem. 2019 Feb 4;13(1):20. doi: 10.1186/s13065-019-0539-1. eCollection 2019 Dec.

Abstract

BACKGROUND

Measuring concentrations of metabolites of estradiol and progesterone in urine, instead of measuring serum concentrations, is common in research and also is used in patient care. The primary aim of this study was to demonstrate that analysis of urine samples dried on filter paper by gas chromatography with tandem mass spectrometry (GC-MS/MS) provides results similar to serum analyzed by radioimmunoassay (RIA). Secondary aims were to show that collection of four samples during the day (4-spot method) can be substituted for a 24-h collection, and that analysis of urine from dried samples is equivalent to liquid urine samples.

METHODS

This prospective observational study compared results of urine and serum analyses. Urine samples from women throughout the menstrual cycle and single samples from postmenopausal women were evaluated. Urine was collected onto filter paper and dried. Dried urine was extracted, hydrolyzed, and derivatized prior to analysis by GC-MS/MS. Hormone concentrations were normalized to creatinine. Single samples were used to compare results of 24-h urine collection to the 4-spot method from a separate population of women and men. A subset of these samples were used to compare results from dried urine to liquid urine.

RESULTS

The primary study showed good reliability in the comparisons between the dried urine and serum assays. During the menstrual cycles of a subset of four women, urine metabolite concentrations followed the same pattern as serum concentrations. Comparison of 4-spot to 24-h urine collections and of dried to liquid urine measurements had intraclass correlation coefficients (ICC) greater than 0.95, indicating excellent agreement.

CONCLUSIONS

For estradiol and progesterone, the dried urine assay is a good surrogate for serum testing. The 4-spot method can be used instead of 24-h urine collections and dried urine results are comparable to liquid urine. The dried urine assay is useful for some clinical assessments of hormone disorders and may be useful in large epidemiologic studies due to ease of sample handling.

摘要

背景

在研究中,测定尿液中雌二醇和孕酮的代谢物浓度而非血清浓度很常见,并且也用于患者护理。本研究的主要目的是证明通过气相色谱 - 串联质谱法(GC-MS/MS)分析滤纸上干燥的尿液样本所得到的结果与通过放射免疫分析法(RIA)分析血清的结果相似。次要目的是表明日间采集四个样本(四点法)可替代24小时采集,以及干燥样本的尿液分析等同于液体尿液样本分析。

方法

这项前瞻性观察性研究比较了尿液和血清分析结果。评估了整个月经周期女性的尿液样本以及绝经后女性的单个样本。尿液收集到滤纸上并干燥。干燥的尿液在通过GC-MS/MS分析之前进行提取、水解和衍生化。激素浓度以肌酐进行标准化。使用单个样本将24小时尿液收集的结果与来自另一组女性和男性的四点法结果进行比较。这些样本的一个子集用于比较干燥尿液和液体尿液的结果。

结果

主要研究表明干燥尿液检测与血清检测之间的比较具有良好的可靠性。在四名女性的一个子集中的月经周期期间,尿液代谢物浓度与血清浓度遵循相同模式。四点法与24小时尿液收集以及干燥尿液与液体尿液测量的组内相关系数(ICC)大于0.95,表明一致性极佳。

结论

对于雌二醇和孕酮,干燥尿液检测是血清检测的良好替代方法。可以使用四点法代替24小时尿液收集,并且干燥尿液结果与液体尿液相当。干燥尿液检测对于激素紊乱的一些临床评估有用,并且由于样本处理简便,可能在大型流行病学研究中有用。