Garcerant Daniel, Jiménez-Alfaro Ignacio, Alejandre Nicolás
Fundación Jiménez Díaz, Madrid, Spain.
J Ophthalmol. 2019 Jul 18;2019:2593404. doi: 10.1155/2019/2593404. eCollection 2019.
To establish which reference body offers the greatest sensitivity in keratoconus (KC) diagnosis, obtain normative data for the myopic population with toric ellipsoid reference bodies, and determine the cutoff points for a population with KC.
A retrospective, observational study of the entire Scheimpflug tomographer database of the Fundación Jiménez Díaz in Madrid was conducted to identify a normal myopic and a KC myopic population. Three different reference bodies were tested on all patients: best fit sphere (BFS), best fit toric ellipsoid with fixed eccentricity (BFTEFE), and best fit toric ellipsoid (BFTE). Anterior and posterior elevation measurements at the apex and thinnest point were recorded, as well as the root mean square of posterior elevations (RMS-P). Normative data were extracted, and receiver operating characteristic (ROC) curves were generated to obtain cutoff points between the normal and KC population.
A total of 301 eyes were included, comprising 219 normal myopic and 82 myopic KC eyes. BFS and BFTEFE produced the best results when measuring posterior elevation at the thinnest point. BFTE had better sensitivity with the RMS-P. From all measurements, best sensitivity (100%) was achieved with a cutoff point of 8 m of posterior elevation at the thinnest point using the BFTEFE. BFTE was found to hide the cone in certain patients.
Posterior elevation measured at the thinnest point with a BFTEFE is the best-performing parameter and, therefore, is recommended to discriminate between normal and KC patients within a myopic population.
确定哪种参考体在圆锥角膜(KC)诊断中具有最高的敏感性,获取具有环曲面椭球体参考体的近视人群的规范数据,并确定KC人群的截断点。
对马德里希门尼斯·迪亚斯基金会的整个Scheimpflug断层扫描仪数据库进行回顾性观察研究,以识别正常近视人群和KC近视人群。对所有患者测试了三种不同的参考体:最佳拟合球(BFS)、具有固定偏心率的最佳拟合环曲面椭球体(BFTEFE)和最佳拟合环曲面椭球体(BFTE)。记录顶点和最薄点处的前表面和后表面高度测量值,以及后表面高度的均方根(RMS-P)。提取规范数据,并生成受试者操作特征(ROC)曲线以获得正常人群和KC人群之间的截断点。
共纳入301只眼,其中219只正常近视眼和82只KC近视眼。在测量最薄点处的后表面高度时,BFS和BFTEFE产生了最佳结果。BFTE在RMS-P方面具有更好的敏感性。在所有测量中,使用BFTEFE在最薄点处后表面高度截断点为8μm时获得了最佳敏感性(100%)。发现BFTE在某些患者中隐藏了圆锥。
使用BFTEFE在最薄点处测量的后表面高度是表现最佳的参数,因此,建议用于区分近视人群中的正常患者和KC患者。