Identifying lung overdistention during mechanical ventilation by using volume-pressure loops.

作者信息

Fisher J B, Mammel M C, Coleman J M, Bing D R, Boros S J

机构信息

Infant Pulmonary Research Center of Children's Hospital, St. Paul, Minnesota 55102.

出版信息

Pediatr Pulmonol. 1988;5(1):10-4. doi: 10.1002/ppul.1950050104.

Abstract

We measured the pulmonary mechanics of 23 mechanically ventilated neonates. Airway pressures, inspiratory and expiratory flows were simultaneously measured. Values for respiratory system mechanics were then derived from these data by using a personal computer and a special software program. Volume-pressure (V-P) loops and respiratory system compliance values were determined for representative mechanical breaths. Twelve infants had normal-appearing V-P loops. Eleven had V-P loops characteristic of lung overdistention, showing decreasing changes in volume with progressive increases in pressure. To quantify this visual observation, we determined the change in compliance during the last 20% of inspiration (C20). We then compared this value to the total compliance value for the entire breath (C) using the ratio C20/C. Mean values for C, C20, and C20/C were compared for the two patient groups. Total respiratory system compliance values were similar. C20 values were decreased in those patients with V-P loops showing overdistention. C20/C values were significantly decreased in those patients with V-P loop evidence of overdistention. Patients with V-P loop evidence of overdistention all had C20/C values less than 0.8. Those with normal-appearing V-P loops all had C20/C values greater than 1.0. The C20/C ratio appears to effectively quantitate visual V-P loop evidence of lung overdistention during mechanical ventilation.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验